Investigation of Halcyon multi-leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm.
Article in Journal of applied clinical medical physics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 7 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
7 citing papers in PubMed, 13 citations in OpenAlex.
- Spiral volumetric modulated arc therapy enhances peak-to-valley dose contrast for 3D-GRID radiotherapy compared to VMAT and tomotherapy.BMC cancer · 2026Article
- Solution for the hot spots in adaptive radiotherapy planning for cervical cancer: a case report and review of literature.Frontiers in oncology · 2026Article
- Multi-institutional study on the commissioning and clinical implementation of the TrueBeam enhanced leaf model in the Eclipse treatment planning system.Journal of applied clinical medical physics · 2025Article
- Commissioning of Halcyon enhanced leaf model in the Eclipse treatment planning system: Focus on simple slit fields and VMAT dose calculation.Journal of applied clinical medical physics · 2025Article
- Dosimetric comparison of two linear accelerators for lung cancer SBRT using IMRT dose delivery technique: VenusX vs. Edge.European journal of medical research · 2025Article
- Investigation of Halcyon multi-leaf collimator model in Eclipse treatment planning system: A focus on the VMAT dose calculation with the Acuros XB algorithm.Journal of applied clinical medical physics · 2022Article
- DeltaJournal of medical physicsArticle
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors at 2 institutions in 1 country.
Funding
Abstract
purposeThe dual-layer multi-leaf collimator (MLC) in Halcyon involves further complexities in the dose calculation process, because the leaf-tip transmission varies according to the leaf trailing pattern. For the volumetric modulated arc therapy (VMAT) treatment, the prescribed dose for the target volume can be sensitive to the leaf-tip transmission change. This report evaluates the dosimetric consequence due to the uncertainty of the dual-layer MLC model in Eclipse through the dose verifications for clinical VMAT. Additionally, the Halcyon leaf-tip model is empirically adjusted for the VMAT dose calculation with the Acuros XB. MATERIALS AND
methodsFor this evaluation, an in-house program that analyzes the leaf position in each layer was developed. Thirty-two clinical VMAT plans were edited into three leaf sequences: dual layer (original), proximal single layer, or distal single layer. All leaf sequences were verified using Delta4 according to the dose difference (DD) and the global gamma index (GI). To improve the VMAT dose calculation accuracy, the dosimetric leaf gap (DLG) was adjusted to minimize the DD in single-layer leaf sequences.
resultsThe mean of DD were -1.35%, -1.20%, and -1.34% in the dual-layer, proximal single-layer, and distal single-layer leaf sequences, respectively. The changes in the mean of DD between leaf sequences were within 0.2%. However, the calculated doses differed from the measured doses by approximately 1% in all leaf sequences. The tuned DLG was increased by 0.8 mm from the original DLG in Eclipse. When the tuned DLG was used in the dose calculation, the mean of DD neared 0% and GI with a criterion of 2%/2 mm yielded a pass rate of more than 98%.
conclusionNo significant change was confirmed in the dose calculation accuracy between the leaf sequences. Therefore, it is suggested that the dosimetric consequence due to the leaf trailing was negligibly small in clinical VMAT plans. The DLG tuning for Halcyon can be useful for reducing the dose calculation uncertainties in Eclipse VMAT and required in the commissioning for Acuros XB.
Indexed as