Optimization of dual energy computed tomography post-processing to reduce lower limb artifacts in gout.
Article in Quantitative imaging in medicine and surgery, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
6 citing papers in PubMed, 10 citations in OpenAlex.
- Precision imaging biomarkers for gout: artifact-specific spectral CT parameters for personalized differentiation of tophi in a prospective study.Therapeutic advances in musculoskeletal disease · 2026Article
- Deep learning reconstruction enhances tophus detection in a dual-energy CT phantom study.Scientific reports · 2025Article
- Impact of Dual-Energy Computed Tomography (DECT) Postprocessing Protocols on Detection of Monosodium Urate (MSU) Deposits in Foot Tendons of Cadavers.Diagnostics (Basel, Switzerland) · 2023Article
- Dual-energy computed tomography in reducing the effective radiation dose of computed tomography urography in patients with urinary calculi.Quantitative imaging in medicine and surgery · 2023Article
- The Clinical Utility of Dual-Energy Computed Tomography in the Diagnosis of Gout-A Cross-Sectional Study.Journal of clinical medicine · 2022Article
- Prevalence of Monosodium Urate (MSU) Deposits in Cadavers Detected by Dual-Energy Computed Tomography (DECT).Diagnostics (Basel, Switzerland) · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors at 4 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundIn gout, several types of dual-energy computed tomography (DECT) artifacts have been described (nail bed, skin, beam hardening, submillimeter and vascular artifacts), which can lead to overdiagnosis. The objective of this study was to determine the optimal DECT settings for post processing in order to reduce the frequency of some common artifacts in patients with suspected gout.
methodsSeventy-seven patients hospitalized for suspected gout (feet/ankles and/or knees) who received a DECT imaging were included (final diagnosis of 43 gout and 34 other rheumatic disorders). Different post-processing settings were evaluated using Syngovia software: nine settings (R1 to R9) were evaluated with a combination of different ratio (1.28, 1.36 and 1.55) and attenuation coefficient (120, 150, 170 HU).
resultsAmong the nine settings tested, the R2 setting (170 HU, ratio =1.28) significantly reduced the presence of knee and foot/ankle artifacts compared to the standard R1 setting (85% and 94% decrease in beam hardening and clumpy artifacts in the ankle and foot, respectively (P<0.001); a decrease of 71%, 60% and 88% respectively of meniscal beam hardening, beam hardening and submillimeter artifacts in the knee (P<0.001). Compared to standard settings, the use of R2 settings decreased sensitivity [0.79 (95% CI: 0.65, 0.88) versus 0.90 (95% CI: 0.78, 0.96)] and increased specificity [0.86 (95% CI: 0.71, 0.93) versus 0.63 (95% CI: 0.47, 0.77)] (P<0.001). Settings using an attenuation coefficient to 120 HU and/or a ratio to 1.55 were all associated with a significant increasing of artifacts, especially clumpy and beam hardening artifacts.
conclusionsApplying a ratio of 1.28 and a minimum attenuation of 170 HU in DECT post-processing eliminates the majority of artifacts located in the lower limbs, particularly clumpy artifacts and beam hardening.