Comparison of the Agatston score acquired with photon-counting detector CT and energy-integrating detector CT: ex vivo study of cadaveric hearts.

Susann Skoog, Lilian Henriksson, Håkan Gustafsson, Mårten Sandstedt, Sebastian Elvelind, Anders Persson

Open access · hybridAbstract read
In one paragraph

Article in The international journal of cardiovascular imaging, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.

0numbers the graph read from it
0cells of the map it votes in
9citing papers in PubMed
1.5field-weighted citation impact, top 20% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

9 citing papers in PubMed, 19 citations in OpenAlex.

  1. Article
  2. Review
  3. Review
  4. Review
  5. Review
  6. Review
  7. Review
  8. Review
  9. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 1 institution in 1 country.

Susann SkoogDepartment of Radiology and Department of Health, Medicine and Caring Sciences, Linköping University, 581 85, Linköping, Sweden. Susann.Skoog@regionostergotland.se.ORCID http://orcid.org/0000-0001-5309-0829
Lilian HenrikssonDepartment of Radiology and Department of Health, Medicine and Caring Sciences, Linköping University, 581 85, Linköping, Sweden.ORCID http://orcid.org/0000-0001-5485-6769
Håkan GustafssonCenter for Medical Image Science and Visualization (CMIV), Linköping University, Linköping, Sweden.ORCID http://orcid.org/0000-0001-6148-1053
Mårten SandstedtDepartment of Radiology and Department of Health, Medicine and Caring Sciences, Linköping University, 581 85, Linköping, Sweden.ORCID http://orcid.org/0000-0002-0973-9499
Sebastian ElvelindDepartment of Clinical Pathology, and Department of Biomedical and Clinical Sciences, Linköping University, Linköping, Sweden.
Anders PerssonDepartment of Radiology and Department of Health, Medicine and Caring Sciences, Linköping University, 581 85, Linköping, Sweden.
Linköping University · SE

Funding

Linköpings Universitet LIO 899441
6 · The paper itself

Abstract

The purpose of this study was to compare the correlation and agreement between AS derived from either an energy-integrating detector CT (EID-CT) or a photon-counting detector CT (PCD-CT). Reproducibility was also compared. In total, 26 calcified coronary lesions (from five cadaveric hearts) were identified for inclusion. The hearts were positioned in a chest phantom and scanned in both an EID-CT and a prototype PCD-CT. The EID-CT and PCD-CT acquisition and reconstruction parameters were matched. To evaluate the reproducibility, the phantom was manually repositioned, and an additional scan was performed using both methods. The EID-CT reconstructions were performed using the dedicated calcium score kernel Sa36. The PCD-CT reconstructions were performed with a vendor-recommended kernel (Qr36). Several monoenergetic energy levels (50-150 keV) were evaluated to find the closest match with the EID-CT scans. A semi-automatic evaluation of calcium score was performed on a post-processing multimodality workplace. The best match with Sa36 was PCD-CT Qr36 images, at a monoenergetic level of 72 keV. Statistical analyses showed excellent correlation and agreement. The correlation and agreement with regards to the Agatston score (AS) between the two methods, for each position as well as between the two positions for each method, were assessed with the Spearman´s rank correlation. The correlation coefficient, rho, was 0.98 and 0.97 respectively 0.99 and 0.98. The corresponding agreements were investigated by means of Bland-Altman plots. High correlation and agreement was observed between the AS derived from the EID-CT and a PCD-CT. Both methods also demonstrated excellent reproducibility.

Indexed as

ArteriosclerosisCalciumCoronary vesselsCT-angiographyHeartPhoton counting detector CT

Identifiers

PMID34988781
PMCPMC11142966
OpenAlexW4225326149

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.