Evidence map›Paper›PMID 34980022›Full record

ArticleBMC ophthalmology2022

Correlation between intraoperative and postoperative vaulting of the EVO implantable Collamer lens: a retrospective study of real-time observations of vaulting using the RESCAN 700 system.

Nian Guan, Xiao-Nong Zhang, Wan-Jun Zhang

Erratum issuedOpen access · goldAbstract read
In one paragraph

Article in BMC ophthalmology, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. An erratum has been issued. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
1.3field-weighted citation impact, top 22% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 9 citations in OpenAlex.

  1. Article
  2. Observational
  3. Article
  4. Article
4 · The record

Corrections and comments

5 · Who and what money

Authors and funding

3 authors at 1 institution in 1 country.

Nian GuanDepartment of Refractive, Wuhan Bright Eye Hospital, Wuhan, 430000, Hubei, China.
Xiao-Nong ZhangDepartment of Refractive, Hefei Bright Eye Hospital, Hefei, 230000, Anhui, China.
Wan-Jun ZhangDepartment of Refractive, Hefei Bright Eye Hospital, Hefei, 230000, Anhui, China. 8688560@qq.com.
Hefei Infectious Disease Hospital · CN

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundImplantable Collamer lens (ICL) vaulting is one of the most important parameters for the safety, aqueous humor circulation, and lens transparency after ICL implantation. This study aimed to investigate the factors associated with the actual vaulting after refractive EVO-ICL surgery.

methodsThis retrospective study included patients who underwent EVO-ICL surgery at a tertiary eye hospital between October and December 2019. A RESCAN 700 was used for the intraoperative and CIRRUS HD-OCT was used for postoperative observation of vaulting. Subjective and objective refractions, anterior ocular segment, corneal morphology, intraocular pressure (IOP), anterior chamber volume (ACV), crystalline lens rise (CLR), white-to-white distance (WTW), anterior chamber depth (ACD), axial length, corneal endothelial cell density (ECD), and fundoscopy were examined. A multivariable analysis was performed to determine the factors independently associated with 1-month postoperative vaulting.

resultsFifty-one patients (102 eyes) were included. Compared with the eyes with normal vaulting, those with high vaulting had higher preoperative diopter values (P = 0.039), lower preoperative corrected visual acuity (P = 0.006), lower preoperative IOP (P = 0.029), higher preoperative ACD (P = 0.004), lower preoperative CLR (P = 0.046), higher ICL spherical equivalent (P = 0.030), higher intraoperative vaulting (P < 0.001), and lower IOP at 1 month (P = 0.045). The multivariable analysis showed that the only factor independently associated with high vaulting at 1 month after surgery was the intraoperative vaulting value (odds ratio = 1.005, 95% confidence interval: 1.002-1.007, P < 0.001). The intraoperative and 1-month postoperative vaulting values were positively correlated (R

conclusionsThe RESCAN700 system can be used to perform intraoperative optical coherence tomography to predict the vaulting value of ICL at 1 month.

Indexed as

MyopiaPhakic Intraocular LensesHumansLens Implantation, IntraocularRefraction, OcularRetrospective StudiesLenses, intraocularlens implantation, intraocularPost-operative vaultRefractive surgeryTomography, optical coherence

Identifiers

PMID34980022
PMCPMC8721482
OpenAlexW4206043628

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.