Inflow cannula position as risk factor for stroke in patients with HeartMate 3 left ventricular assist devices.

Thomas Schlöglhofer, Philipp Aigner, Marcel Migas, Dietrich Beitzke, Kamen Dimitrov, Franziska Wittmann, Julia Riebandt, Marcus Granegger, Dominik Wiedemann, Günther Laufer and 3 more

Abstract read
In one paragraph

Article in Artificial organs, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 8 papers.

0numbers the graph read from it
0cells of the map it votes in
8citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

8 citing papers in PubMed.

  1. Article
  2. The impact of position changes of the inflow cannula in LVAD patients.Turk gogus kalp damar cerrahisi dergisi · 2026
    Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

13 authors.

Thomas SchlöglhoferDepartment of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.ORCID https://orcid.org/0000-0003-4354-4860
Philipp AignerLudwig Boltzmann Institute for Cardiovascular Research, Vienna, Austria.ORCID https://orcid.org/0000-0002-3212-2112
Marcel MigasCenter for Medical Physics and Biomedical Engineering, Medical University of Vienna, Vienna, Austria.
Dietrich BeitzkeDepartment of Biomedical Imaging and Image-guided Therapy, Medical University of Vienna, Vienna, Austria.
Kamen DimitrovDepartment of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.
Franziska WittmannDepartment of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.
Julia RiebandtDepartment of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.
Marcus GraneggerDepartment of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.
Dominik WiedemannDepartment of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.
Günther LauferDepartment of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.
Francesco MoscatoLudwig Boltzmann Institute for Cardiovascular Research, Vienna, Austria.ORCID https://orcid.org/0000-0003-0279-6615
Heinrich SchimaDepartment of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.
Daniel ZimpferDepartment of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.

Funding

Oesterreichische Nationalbank 17314
6 · The paper itself

Abstract

backgroundA relation between the left ventricular assist device inflow cannula (IC) malposition and pump thrombus has been reported. This study aimed to investigate if the pump position, derived from chest X-rays in HeartMate 3 (HM3) patients, correlates with neurological dysfunction (ND), ischemic stroke (IS), hemorrhagic stroke (HS) and survival.

methodsThis analysis was performed on routinely acquired X-rays of 42 patients implanted with a HM3 between 2014 and 2017. Device position was quantified in patients with and without ND from frontal and lateral X-rays characterizing the IC and pump in relation to spine, diaphragm or horizontal line. The primary end-point was freedom from stroke and survival one-year after HM3 implantation stratified by pump position.

resultsThe analysis of X-rays, 33.5 (41.0) days postoperative, revealed a significant smaller IC angle of HM3 patients with ND versus no ND (0.1° ± 14.0° vs. 12.9° ± 10.1°, p = 0.005). Additionally, the IC angle in the frontal view, IS: 4.1 (20.9)° versus no IS: 13.8 (7.5)°, p = 0.004 was significantly smaller for HM3 patients with IS. Using receiver operating characteristics derived cut-off, IC angle <10° provided 75% sensitivity and 100% specificity (C-statistic = 0.85) for predicting IS. Stratified by IC angle, freedom from IS at 12 months was 100% (>10°) and 60% (<10°) respectively (p = 0.002). No significant differences were found in any end-point between patients with and without HS. One-year survival was significantly higher in patients with IC angle >10° versus <10° (100% vs. 71.8%, p = 0.012).

conclusionsIC malposition derived from standard chest X-rays serves as a risk factor for ND, IS and worse survival in HM3 patients.

Indexed as

Heart-Assist DevicesHeart FailureStrokeThrombosisCannulaHumansRetrospective StudiesRisk Factorsinflow cannula positionmechanical circulatory supportneurological dysfunctionpredictorventricular assist device