ArticleFrontiers in bioengineering and biotechnology2021
Comparative Analysis of the Biomechanical Characteristics After Different Minimally Invasive Surgeries for Cervical Spondylopathy: A Finite Element Analysis.
Article in Frontiers in bioengineering and biotechnology, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. An erratum has been issued. Cited by 10 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
10 citing papers in PubMed, 1 synthesis or guideline pooled it, 12 citations in OpenAlex.
- Pooled it
- Biomechanical effects of a novel anatomic fusion cage for anterior cervical corpectomy: a finite element study.BMC musculoskeletal disorders · 2026Article
- Development and Biomechanical Study of the Novel PRUNUS Plate-Screw System.Orthopaedic surgery · 2026Article
- Biomechanical effect of cage size in single-level anterior cervical discectomy and fusion: a finite element analysis.BMC musculoskeletal disorders · 2025Article
- Biomechanical analysis of a newly designed and 3D printed plate-locking interbody cage: an observational study of finite element analysis.Scientific reports · 2025Observational
- A novel posterior endoscopic cervical approach for treating cervical spondylotic radiculopathy: a finite-element analysis(C2-T1).Frontiers in bioengineering and biotechnology · 2025Article
- Finite element analysis of a new preoperative traction for cervical kyphosis: suspensory traction.Medical & biological engineering & computing · 2024Article
- Anterior percutaneous full-endoscopic transcorporeal decompression for cervical disc herniation: a finite element analysis and long-term follow-up study.BMC musculoskeletal disorders · 2024Article
- Anterior percutaneous full-endoscopic transcorporeal decompression of the spinal cord via one vertebra with two bony channels for adjacent two-segment cervical spondylotic myelopathy: a technical note.BMC musculoskeletal disorders · 2023Article
- Biomechanical performance of the novel assembled uncovertebral joint fusion cage in single-level anterior cervical discectomy and fusion: A finite element analysis.Frontiers in bioengineering and biotechnology · 2023Article
Corrections and comments
- Erratum issued
Authors and funding
13 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
Minimally invasive surgeries, including posterior endoscopic cervical foraminotomy (PECF), microsurgical anterior cervical foraminotomy (MACF), anterior transdiscal approach of endoscopic cervical discectomy (ATd-ECD), and anterior transcorporeal approach of endoscopic cervical discectomy (ATc-ECD), have obtained positive results for cervical spondylotic radiculopathy. Nonetheless, there is a lack of comparison among them regarding their biomechanical performance. The purpose of this study is to investigate the biomechanical changes of operated and adjacent segments after minimally invasive surgeries compared to a normal cervical spine. A three-dimensional model of normal cervical vertebrae C3-C7 was established using finite element analysis. Afterwards, four surgical models (PECF, MACF, ATd-ECD, and ATc-ECD) were constructed on the basis of the normal model. Identical load conditions were applied to simulate flexion, extension, lateral bending, and axial rotation of the cervical spine. We calculated the range of motion (ROM), intradiscal pressure (IDP), annulus fibrosus pressure (AFP), uncovertebral joints contact pressure (CPRESS), and facet joints CPRESS under different motions. For all circumstances, ATc-ECD was close to the normal cervical spine model, whereas ATd-ECD significantly increased ROM and joints CPRESS and decreased IDP in the operated segment. PECF increased more the operated segment ROM than did the MACF, but the MACF obtained maximum IDP and AFP. Except for ATc-ECD, the other models increased joints CPRESS of the operated segment. For adjacent segments, ROM, IDP, and joints CPRESS showed a downward trend in all models. All models showed good biomechanical stability. With their combination biomechanics, safety, and conditions of application, PECF and ATc-ECD could be appropriate choices for cervical spondylotic radiculopathy.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.