ArticleJournal of applied clinical medical physics2022

Dosimetric evaluation of irradiation geometry and potential air gaps in an acrylic miniphantom used for external audit of absolute dose calibration for a hybrid 1.5 T MR-linac system.

Neelam Tyagi, Ergys Subashi, Dale Michael Lovelock, Stephen Kry, Paola Elisa Alvarez, Margie A Hunt, Seng Boh Lim

Open access · goldAbstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
1.6field-weighted citation impact, top 16% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 8 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 2 institutions in 1 country.

Neelam TyagiDepartment of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York, New York, USA.
Ergys SubashiDepartment of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York, New York, USA.
Dale Michael LovelockDepartment of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York, New York, USA.
Stephen KryDepartment of Radiation Physics, IROC, MD Anderson Cancer Center, Houston, Texas, USA.
Paola Elisa AlvarezDepartment of Radiation Physics, IROC, MD Anderson Cancer Center, Houston, Texas, USA.
Margie A HuntDepartment of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York, New York, USA.
Seng Boh LimDepartment of Medical Physics, Memorial Sloan-Kettering Cancer Center, New York, New York, USA.
Memorial Sloan Kettering Cancer Center · USThe University of Texas MD Anderson Cancer Center · US

Funding

X-RAY CRYSTALLOGRAPHYP30CA008748 · NCI · SLOAN-KETTERING INSTITUTE FOR CANCER RES · PI Elisa de Stanchina · 1985 to 2026
$347.4M
IROC: Enhancing local enrolling site radiological data capture capabilities for NCTN trialsU24CA180803 · NCI · AMERICAN COLLEGE OF RADIOLOGY · PI Thomas J. FitzGerald, MICHAEL V KNOPP · 2014 to 2026
$116.2M
NCI NIH HHS P30 CA008748NCI NIH HHS U24 CA180803Support Grant CA180803
6 · The paper itself

Abstract

introductionTo investigate the impact of partial lateral scatter (LS), backscatter (BS) and presence of air gaps on optically stimulated luminescence dosimeter (OSLD) measurements in an acrylic miniphantom used for dosimetry audit on the 1.5 T magnetic resonance-linear accelerator (MR-linac) system.

methodsThe following irradiation geometries were investigated using OSLDs, A26 MR/A12 MR ion chamber (IC), and Monaco Monte Carlo system: (a) IC/OSLD in an acrylic miniphantom (partial LS, partial BS), (b) IC/OSLD in a miniphantom placed on a solid water (SW) stack at a depth of 1.5 cm (partial LS, full BS), (c) IC/OSLD placed at a depth of 1.5 cm inside a 3 cm slab of SW/buildup (full LS, partial BS), and (d) IC/OSLD centered inside a 3 cm slab of SW/buildup at a depth of 1.5 cm placed on top of a SW stack (full LS, full BS). Average of two irradiated OSLDs with and without water was used at each setup. An air gap of 1 and 2 mm, mimicking presence of potential air gap around the OSLDs in the miniphantom geometry was also simulated. The calibration condition of the machine was 1 cGy/MU at SAD = 143.5 cm, d = 5 cm, G90, and 10 × 10 cm

resultsThe Monaco calculation (0.5% uncertainty and 1.0 mm voxel size) for the four setups at the measurement point were 108.2, 108.1, 109.4, and 110.0 cGy. The corresponding IC measurements were 109.0 ± 0.03, 109.5 ± 0.06, 110.2 ± 0.02, and 109.8 ± 0.03 cGy. Without water, OSLDs measurements were ∼10% higher than the expected. With added water to minimize air gaps, the measurements were significantly improved to within 2.2%. The dosimetric impacts of 1 and 2 mm air gaps were also verified with Monaco to be 13.3% and 27.9% higher, respectively, due to the electron return effect.

conclusionsA minimal amount of air around or within the OSLDs can cause measurement discrepancies of 10% or higher when placed in a high b-field MR-linac system. Care must be taken to eliminate the air from within and around the OSLD.

Indexed as

Particle AcceleratorsRadiometryCalibrationHumansMonte Carlo MethodPhantoms, Imaging1.5 T magnetic fieldelectron return effect (ERE)IROCMR-guided radiation therapyOSL

Identifiers

PMID34914175
PMCPMC8833292
OpenAlexW4200124366

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.