ArticleBMC anesthesiology2021
Preparation of Dräger Atlan A350 and General Electric Healthcare Carestation 650 anesthesia workstations for malignant hyperthermia susceptible patients.
Article in BMC anesthesiology, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed, 4 citations in OpenAlex.
- Desflurane Safety Revisited: A Pharmacovigilance Study Detecting Potential Safety Signals from FAERS Data.Journal of pain research · 2025Article
- Rebounds of sevoflurane concentration during simulated trigger-free pediatric and adult anesthesia.BMC anesthesiology · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
8 authors at 3 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundPatients at risk of malignant hyperthermia need trigger-free anesthesia. Therefore, anesthesia machines prepared for safe use in predisposed patients should be free of volatile anesthetics. The washout time depends on the composition of rubber and plastic in the anesthesia machine. Therefore, new anesthesia machines should be evaluated regarding the safe preparation for trigger-free anesthesia. This study investigates wash out procedures of volatile anesthetics for two new anesthetic workstations: Dräger Atlan A350 and General Electric Healthcare (GE) Carestation 650 and compare it with preparation using activated charcoal filters (ACF).
methodsA Dräger Atlan and a Carestation 650 were contaminated with 4% sevoflurane for 90 min. The machines were decontaminated with method (M1): using ACF, method 2 (M2): a wash out method that included exchange of internal parts, breathing circuits and soda lime canister followed by ventilating a test lung using a preliminary protocol provided by Dräger or method 3 (M3): a universal wash out instruction of GE, method 4 (M4): M3 plus exchange of breathing system and bellows. Decontamination was followed by a simulated trigger-free ventilation. All experiments were repeated with 8% desflurane contaminated machines. Volatile anesthetics were detected with a closed gas loop high-resolution ion mobility spectrometer with gas chromatographic pre-separation attached to the bacterial filter of the breathing circuits. Primary outcome was time until < 5 ppm of volatile anesthetics and total preparation time.
resultsTime to < 5 ppm for the Atlan was 17 min (desflurane) and 50 min (sevoflurane), wash out continued for a total of 60 min according to protocol resulting in a total preparation time of 96-122 min. The Carestation needed 66 min (desflurane) and 24 min (sevoflurane) which could be abbreviated to 24 min (desflurane) if breathing system and bellows were changed. Total preparation time was 30-73 min. When using active charcoal filters time to < 5 ppm was 0 min for both machines, and total preparation time < 5 min.
conclusionBoth wash out protocols resulted in a significant reduction of trace gas concentrations. However, due to the complexity of the protocols and prolonged total preparation time, feasibility in clinical practice remains questionable. Especially when time is limited preparation of the anesthetic machines using ACF remain superior.