ArticleSAGE open medicine2021

Comparative structural analysis of polyurethane and silicone catheters of totally implantable venous access devices by micro-computed tomography.

Daniele Panetta, Lara Tollapi, Adriana Paolicchi, Elisa Marri, Giacomo Aringhieri, Davide Caramella

Open access · goldAbstract read
In one paragraph

Article in SAGE open medicine, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
0.6field-weighted citation impact, top 28% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 3 citations in OpenAlex.

  1. Article
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors at 3 institutions in 1 country.

Daniele PanettaCNR Institute of Clinical Physiology, Pisa, Italy.ORCID https://orcid.org/0000-0002-4184-0175
Lara TollapiAnesthesiology and Pain Therapy Unit - Vascular Access Team, University Hospital of Pisa, Pisa, Italy.
Adriana PaolicchiAnesthesiology and Pain Therapy Unit - Vascular Access Team, University Hospital of Pisa, Pisa, Italy.
Elisa MarriDepartment of Translational Research and New Technologies in Medicine and Surgery, University of Pisa, Pisa, Italy.
Giacomo AringhieriDepartment of Translational Research and New Technologies in Medicine and Surgery, University of Pisa, Pisa, Italy.ORCID https://orcid.org/0000-0003-0842-5372
Davide CaramellaDepartment of Translational Research and New Technologies in Medicine and Surgery, University of Pisa, Pisa, Italy.ORCID https://orcid.org/0000-0002-9951-2916
University of Pisa · ITAzienda Ospedaliera Universitaria Pisana · ITIstituto di Fisiologia Clinica · IT

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectivesTo investigate microstructural alterations of explanted long-term central venous catheters of totally implantable venous access devices, using micro-computed tomography.

methodsA total of 16 catheters (9 made of silicone and 7 made of polyurethane), all non-fractured, have been analyzed in this study. Eight catheters were implanted for an average duration of 994 days (min-max: 98-2731 days), while the remaining eight catheters (four for each material, forming the SIref and PUref control groups) were analyzed before implant and used as a reference. X-ray micro-computed tomography was used to reconstruct the three-dimensional geometry of selected segments of each catheter (ca. 10 cm per sample).

resultsMorphometric analysis of the catheters revealed increases wall thickness and section area in the polyurethane group as compared with the reference central venous catheters of the same materials (wall thickness: 403 ± 12 μm in the polyurethane (PU) group vs 382 ± 4 μm in PUref, p = 0.014; wall cross-section area: 2.04 ± 0.09 mm

conclusionMorphological change and local structural alteration can occur in both silicone and polyurethane catheters. This evidence suggests the need for further studies connecting those morphological changes with modification of mechanical robustness, which ultimately can play a role for patient safety.

Indexed as

Central venous cathetersmedical devicesmicro-computed tomographypatient safetystructural alterations

Identifiers

PMID34881041
PMCPMC8646823
OpenAlexW4205963719

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.