Fabrication and evaluation of bilateral Helmholtz radiofrequency coil for thermo-stable breast image with reduced artifacts.
Article in Journal of applied clinical medical physics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Fabrication and evaluation of bilateral Helmholtz radiofrequency coil for thermo-stable breast image with reduced artifacts.Journal of applied clinical medical physics · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
10 authors.
Funding
Abstract
purposeThe positron emission tomography (PET)-magnetic resonance (MR) system is a newly emerging technique that yields hybrid images with high-resolution anatomical and metabolic information. With PET-MR imaging, a definitive diagnosis of breast abnormalities will be possible with high spatial accuracy and images will be acquired for the optimal fusion of anatomic locations. Therefore, we propose a PET-compatible two-channel breast MR coil with minimal disturbance to image acquisition which can be used for simultaneous PET-MR imaging in patients with breast cancer. MATERIALS AND
methodsFor coil design and construction, the conductor loops of the Helmholtz coil were tuned, matched, and subdivided with nonmagnetic components. Element values were optimized with an electromagnetic field simulation. Images were acquired on a GE 600 PET-computed tomography (CT) and GE 3.0 T MR system. For this study, we used the T1-weighted image (volunteer; repetition time (TR), 694 ms; echo time (TE), 9.6 ms) and T2-weighted image (phantom; TR, 8742 ms; TE, 104 ms) with the fast spin-echo sequence.
resultsThe results of measuring image factors with the proposed radiofrequency (RF) coil and standard conventional RF coil were as follows: signal-to-noise ratio (breast; 207.7 vs. 175.2), percent image uniformity (phantom; 89.22%-91.27% vs. 94.63%-94.77%), and Hounsfield units (phantom; -4.51 vs. 2.38).
conclusionsOur study focused on the feasibility of proposed two-channel Helmholtz loops (by minimizing metallic components and soldering) for PET-MR imaging and found the comparable image quality to the standard conventional coil. We believe our work will help significantly to improve image quality with the development of a less metallic breast MR coil.