Evidence map›Paper›PMID 34846588›Full record

ArticleArchives of orthopaedic and trauma surgery2022

Varus malalignment of short femoral stem not associated with post-hip arthroplasty fracture.

Scott T Nishioka, Samantha N Andrews, Kristin Mathews, Cass K Nakasone

Abstract read
PubMed Publisher
In one paragraph

Article in Archives of orthopaedic and trauma surgery, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 12 papers.

0numbers the graph read from it
0cells of the map it votes in
12citing papers in PubMed
2.1field-weighted citation impact, top 12% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

12 citing papers in PubMed, 17 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Observational
  10. Article
  11. Standard versus short stem cemented ExeterArthroplasty (London, England) · 2023
    Article
  12. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors at 2 institutions in 1 country.

Scott T NishiokaBone & Joint Center, Straub Medical Center, 888 South King Street, Honolulu, HI, 96818, USA.
Samantha N AndrewsBone & Joint Center, Straub Medical Center, 888 South King Street, Honolulu, HI, 96818, USA. samantha.andrews@straub.net.ORCID http://orcid.org/0000-0003-1378-7845
Kristin MathewsBone & Joint Center, Straub Medical Center, 888 South King Street, Honolulu, HI, 96818, USA.
Cass K NakasoneBone & Joint Center, Straub Medical Center, 888 South King Street, Honolulu, HI, 96818, USA.
Straub Medical Center · USUniversity of Hawaiʻi at Mānoa · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionPeriprosthetic femoral fractures are an increasingly common post-operative complication of total hip arthroplasty (THA). Though varus malalignment is known to increase fracture risk in standard-length femoral stems, varus malalignment is not as well studied in short stems. Therefore, the purpose of this study was to determine if varus malalignment contributes to early periprosthetic fracture risk in a cementless taper-wedged, short femoral stem. MATERIALS AND

methodsThis retrospective review included 366 consecutive patients (441 THAs) having undergone THA via anterior approach by a single surgeon between July 2014 and December 2016. All patients received the same short, cementless femoral stem. Femoral component angle was measured on 6-week post-THA weight-bearing radiographs, with malalignment defined as a femoral component angle exceeding 0° ± 3°. Periprosthetic femoral fracture and aseptic loosening occurring within 2 years post-THA were recorded.

resultsThe final data analysis included 426 hips with a mean follow-up time of 32.9 ± 10.2 months. Varus and neutral alignment occurred in 84 (19.6%) and 342 (79.9%) of stems, respectively. Three (0.7%) periprosthetic femoral fractures occurred within 2 years, all occurring in patients with neutrally aligned femoral stems. One (0.2%) stem failed due to aseptic loosening and was malaligned.

conclusionDespite nearly 20% of stems placed in varus alignment, three of the four early complications occurred in a neutrally aligned stem. Based on these results, forceful intraoperative realignment of a short femoral stem with good initial fixation may present an unnecessary increased risk of intraoperative fracture.

Indexed as

Arthroplasty, Replacement, HipFemoral FracturesHip FracturesHip ProsthesisPeriprosthetic FracturesHumansProsthesis DesignReoperationRetrospective StudiesDirect anterior approachEarly failureMalalignmentPeriprosthetic fractureShort stem

Identifiers

PMID34846588
OpenAlexW3216183806

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.