Disposable point-of-care portable perfusion phantom for quantitative DCE-MRI.
Article in Medical physics, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 9 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
9 citing papers in PubMed, 11 citations in OpenAlex.
- Quantitative dynamic contrast-enhanced magnetic resonance imaging for early assessment of treatment response in pancreatic ductal adenocarcinoma.European radiology · 2026Article
- Quantitative dynamic contrast-enhanced magnetic resonance imaging for renal perfusion measurement in autosomal dominant polycystic kidney disease.Quantitative imaging in medicine and surgery · 2026Article
- ProAgio, a Novel Integrin αvβ3 Targeted Cytotoxin, Suppresses Tumor Growth and Reprograms the PDAC Microenvironment.bioRxiv : the preprint server for biology · 2026Article
- 3D-printable phantoms for quantitative dynamic contrast-enhanced MRI.Magnetic resonance in medicine · 2025Article
- A correction for modeling radial, spiral, and PROPELLER dynamic contrast-enhanced data: Time-averaged extended Tofts.Magnetic resonance in medicine · 2025Article
- Detecting pseudo versus true progression of glioblastoma via accurate quantitative DCE-MRI using point-of-care portable perfusion phantoms: a pilot study.Quantitative imaging in medicine and surgery · 2025Article
- A signal model for fat-suppressed TNMR in biomedicine · 2025Article
- Phantoms for Quantitative Body MRI: a review and discussion of the phantom value.Magma (New York, N.Y.) · 2024Review
- Perfusion Change of Hepatocellular Carcinoma During Atezolizumab plus Bevacizumab Treatment: A Pilot Study.Journal of gastrointestinal cancer · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
14 authors at 3 institutions in 1 country.
Funding
Abstract
purposeTo develop a disposable point-of-care portable perfusion phantom (DP4) and validate its clinical utility in a multi-institutional setting for quantitative dynamic contrast-enhanced magnetic resonance imaging (qDCE-MRI).
methodsThe DP4 phantom was designed for single-use and imaged concurrently with a human subject so that the phantom data can be utilized as the reference to detect errors in qDCE-MRI measurement of human tissues. The change of contrast-agent concentration in the phantom was measured using liquid chromatography-mass spectrometry. The repeatability of the contrast enhancement curve (CEC) was assessed with five phantoms in a single MRI scanner. Five healthy human subjects were recruited to evaluate the reproducibility of qDCE-MRI measurements. Each subject was imaged concurrently with the DP4 phantom at two institutes using three 3T MRI scanners from three different vendors. Pharmacokinetic (PK) parameters in the regions of liver, spleen, pancreas, and paravertebral muscle were calculated based on the Tofts model (TM), extended Tofts model (ETM), and shutter speed model (SSM). The reproducibility of each PK parameter over three measurements was evaluated with the intraclass correlation coefficient (ICC) and compared before and after DP4-based error correction.
resultsThe contrast-agent concentration in the DP4 phantom was linearly increased over 10 min (0.17 mM/min, measurement accuracy: 96%) after injecting gadoteridol (100 mM) at a constant rate (0.24 ml/s, 4 ml). The repeatability of the CEC within the phantom was 0.997 when assessed by the ICC. The reproducibility of the volume transfer constant, K
conclusionsThe DP4 phantom is reliable, portable, and capable of significantly improving the reproducibility of qDCE-MRI measurements.