Comparison of sliding window and field-in-field techniques for tangential whole breast irradiation using the Halcyon and Synergy Agility systems.

Anne Richter, Sonja Wegener, Kathrin Breuer, Gary Razinskas, Stefan Weick, Florian Exner, Klaus Bratengeier, Michael Flentje, Otto Sauer, Bülent Polat

Open access · goldAbstract readComparative Study
In one paragraph

Article in Radiation oncology (London, England), 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
1.6field-weighted citation impact, top 17% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 6 citations in OpenAlex.

  1. Trial
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

10 authors at 1 institution in 1 country.

Anne RichterDepartment of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany. richter_a3@ukw.de.ORCID http://orcid.org/0000-0002-0598-5641
Sonja WegenerDepartment of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
Kathrin BreuerDepartment of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
Gary RazinskasDepartment of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
Stefan WeickDepartment of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
Florian ExnerDepartment of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
Klaus BratengeierDepartment of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
Michael FlentjeDepartment of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
Otto SauerDepartment of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
Bülent PolatDepartment of Radiation Oncology, University of Wuerzburg, Wuerzburg, Germany.
University of Würzburg · DE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundTo implement a tangential treatment technique for whole breast irradiation using the Varian Halcyon and to compare it with Elekta Synergy Agility plans.

methodsFor 20 patients two comparable treatment plans with respect to dose coverage and normal tissue sparing were generated. Tangential field-in-field treatment plans (Pinnacle/Synergy) were replanned using the sliding window technique (Eclipse/Halcyon). Plan specific QA was performed using the portal Dosimetry and the ArcCHECK phantom. Imaging and treatment dose were evaluated for treatment delivery on both systems using a modified CIRS Phantom.

resultsThe mean number of monitor units for a fraction dose of 2.67 Gy was 515 MUs and 260 MUs for Halcyon and Synergy Agility plans, respectively. The homogeneity index and dose coverage were similar for both treatment units. The plan specific QA showed good agreement between measured and calculated plans. All Halcyon plans passed portal dosimetry QA (3%/2 mm) with 100% points passing and ArcCheck QA (3%/2 mm) with 99.5%. Measurement of the cumulated treatment and imaging dose with the CIRS phantom resulted in lower dose to the contralateral breast for the Halcyon plans.

conclusionsFor the Varian Halcyon a plan quality similar to the Elekta Synergy device was achieved. For the Halcyon plans the dose contribution from the treatment fields to the contralateral breast was even lower due to less interleaf transmission of the Halcyon MLC and a lower contribution of scattered dose from the collimator system.

Indexed as

Phantoms, ImagingFemaleHumansImage Processing, Computer-AssistedNeoplasmsOrgans at RiskParticle AcceleratorsQuality Assurance, Health CareRadiotherapy DosageRadiotherapy, Intensity-ModulatedRadiotherapy Planning, Computer-AssistedRetrospective StudiesTomography, X-Ray ComputedDose to OARsHalcyonIGRTWhole breast irradiation

Identifiers

PMID34742291
PMCPMC8571882
OpenAlexW3196097587

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.