Evaluation and comparison of performance of low-dose 128-slice CT scanner with different mAs values: A phantom study.
Article in Journal of carcinogenesis, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
- Utility of Pulmonary Angiography by 128-Slice Computed Tomographic Scanner in Diagnosis of Tetralogy of Fallot Cases.Case reports in radiology · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
2 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectiveRadiation dose in computed tomography (CT) has been the concern of physicists ever since the introduction of CT scan. The objective of this study was to evaluate the performance of low-dose 128-slice CT scanner with different mAs values. MATERIALS AND
methodsQuantitative study was carried out at different values of mAs. Philips brilliance CT phantom with Philips ingenuity 128-slice low-dose CT scanner was chosen for this study. CT number linearity, CT number accuracy, slice thickness accuracy, high-contrast resolution, and low-contrast resolution were calculated and estimated computed tomography dose index volume (CTDI
resultsData analysis shows that image quality was acceptable for all protocols. High-contrast resolution for all protocols was 20 line pairs per centimeter. Low-contrast resolution for 50 mAs images was 4 mm and 3 mm for other mAs protocols. Images acquired using 100 mAs revealed ring artifacts. CTDI
conclusionIt is evident here that mAs has direct impact on the radiation dose to patient. With iDose4, mAs can be reduced to 50 mAs in multislice low-dose CT scan to reduce the radiation dose with minimal effect on image quality for slice thickness 4 mm. However, noise would dominate at tube current lower than 50 mAs for 120 kVp.