ArticleRespiratory care2022
Reliable Estimates of Power Delivery During Mechanical Ventilation Utilizing Easily Obtained Bedside Parameters.
Article in Respiratory care, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed, 6 citations in OpenAlex.
- The Impact of Mechanical Energy Assessment on Mechanical Ventilation: A Comprehensive Review and Practical Application.Medical science monitor : international medical journal of experimental and clinical research · 2023Review
- The Evolution of Intermittent Mandatory Ventilation.Respiratory care · 2023Review
- Outcomes and risk factors for delayed-onset postoperative respiratory failure: a multi-center case-control study by the University of California Critical Care Research Collaborative (UCBMC anesthesiology · 2022Article
Corrections and comments
- Commented on by
Authors and funding
4 authors at 2 institutions in 1 country.
Funding
Abstract
backgroundVentilator-induced lung injury (VILI) requires repetitive transfer of energy from the ventilator to the compromised lung. To understand this phenomenon, 2 sets of equations have been developed to partition total inflation energy into harmless and hazardous components using an arbitrary level of alveolar pressure as a threshold beyond which further energy increments may become damaging. One set of equations uses premeasured resistance and compliance as inputs to predict the energy that would be delivered by typical ventilator settings, whereas the other equation set uses observed output values for end-inspiratory peak and plateau pressure of an already completed inflation.
methodsOur aim for this study was to compare the relative accuracy of these equation sets against the performance of a physical one-compartment model of the respiratory system, programmed with information readily available at the bedside and ventilated using both constant and decelerating flow profiles. Accordingly, equations of each set were compared against the corresponding energy areas measured by digital planimetry of pressure-volume curves for 76 ventilator and patient parameter combinations and over 500 power calculations.
resultsWith few exceptions, all equations strongly correlated with their corresponding measurements by planimetry.
conclusionsThis validation of threshold-partitioned energy equations suggests their potential utility for implementing practical strategies for VILI avoidance.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.