Evidence map›Paper›PMID 34635754›Full record

ArticleScientific reports2021

Thermal effect of a 445 nm diode laser on five dental implant systems: an in vitro study.

Herbert Deppe, Markus Ahrens, Alexandra V Behr, Christina Marr, Anton Sculean, Petra Mela, Lucas M Ritschl

Open access · goldAbstract read
In one paragraph

Article in Scientific reports, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 12 papers.

0numbers the graph read from it
0cells of the map it votes in
12citing papers in PubMed
3.5field-weighted citation impact, top 7% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

12 citing papers in PubMed, 24 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Article
  11. Article
  12. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 2 institutions in 2 countries.

Herbert DeppeDepartment of Oral and Maxillofacial Surgery, School of Medicine, Technical University of Munich, Ismaninger Straße 22, 81675, Munich, Germany.
Markus AhrensDepartment of Mechanical Engineering and Munich School of BioEngineering, Chair of Medical Materials and Implants, Technical University of Munich, Boltzmannstraße 15, 85748, Garching, Germany.
Alexandra V BehrDepartment of Oral and Maxillofacial Surgery, School of Medicine, Technical University of Munich, Ismaninger Straße 22, 81675, Munich, Germany.
Christina MarrDepartment of Oral and Maxillofacial Surgery, School of Medicine, Technical University of Munich, Ismaninger Straße 22, 81675, Munich, Germany.
Anton SculeanDepartment of Periodontology, University of Berne, Freiburgstraße 7, 3010, Berne, Switzerland.
Petra Mela *Department of Mechanical Engineering and Munich School of BioEngineering, Chair of Medical Materials and Implants, Technical University of Munich, Boltzmannstraße 15, 85748, Garching, Germany.
Lucas M Ritschl *Department of Oral and Maxillofacial Surgery, School of Medicine, Technical University of Munich, Ismaninger Straße 22, 81675, Munich, Germany. lucas.ritschl@tum.de.
Technical University of Munich · DEUniversity of Bern · CH

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

The purpose of this in vitro study was to assess the thermal effect of the 445 nm diode laser on five dental implant systems. In an ailing implant protocol, five commercial dental implant systems were subjected to 445 nm diode laser energy at different wattages [W], exposure times, and modes (continuous wave [CW] vs. pulsed and contact vs. non-contact) of laser beam delivery. Scanning electron microscopy (SEM) allowed the evaluation of irradiated implant surfaces. A total of 2880 temperature response curves were recorded. The 445 nm wavelength caused temperature increases of more than 10 °C at or above the 0.8 W power level working in CW mode for 5 s and in pulsed mode at 3 W for 20 s with 10% duty cycle. Highest rises in temperature were seen in the Straumann Pure ceramic implant, lowest in the Ankylos system. SEM analysis revealed no surface alteration in all systems in non-contact mode. The applied laser is not inherently safe for the decontamination of ailing implants. From the results of this study it was concluded that different dental implant materials and geometries show different temperature response curves when subjected to 445 nm diode laser energy. Clinicians ought to be aware of this. Therefore, manufacturers of laser devices should provide implant-specific laser parameters for the decontamination process. However, both laser irradiation systems can prevent harmful rises in temperature and surface alteration when used at moderate laser parameters.

Indexed as

TemperatureDecontaminationDental ImplantsHumansIn Vitro TechniquesLasers, SemiconductorMicroscopy, Electron, ScanningSurface PropertiesTitaniumDental ImplantsTitanium

Identifiers

PMID34635754
PMCPMC8505640
OpenAlexW3207348591

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.