Evaluation of the Elecsys SARS-CoV-2 antigen assay for the detection of SARS-CoV-2 in nasopharyngeal swabs.

Myriam Ben Abdelhanin, Dieudonné M Mvumbi, Aurélie Agathine, Nikolaos Nanos, Stéphane Gidenne

Open access · greenAbstract read
In one paragraph

Article in Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed, 1 pooled it
1.0field-weighted citation impact, top 21% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 1 synthesis or guideline pooled it, 13 citations in OpenAlex.

  1. Pooled it
  2. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors at 3 institutions in 1 country.

Myriam Ben AbdelhaninDepartment of Clinical Chemistry, LBS Laboratory, Cerba HealthCare, Brussels, Belgium. Electronic address: mbenhanin@outlook.be.
Dieudonné M MvumbiDepartment of Expertise and Service provision, Quality of Laboratories, Sciensano, Brussels, Belgium.
Aurélie AgathineDepartment of Clinical Microbiology, LBS Laboratory, Cerba HealthCare, Brussels, Belgium.
Nikolaos NanosDepartment of Clinical Serology, LBS Laboratory, Cerba HealthCare, Brussels, Belgium.
Stéphane GidenneDepartment of Clinical Microbiology, Ketterthill Laboratory, Cerba HealthCare, Belvaux, Luxembourg.
Agfa HealthCare · BEBesins Healthcare (Belgium) · BESciensano (Belgium) · BE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

aimsThe aim of this study was to evaluate the performance of the automated Elecsys® SARS-CoV-2 antigen assay compared to RT-PCR taken as the gold standard for SARS-CoV-2 detection.

methods225 nasopharyngeal swabs were randomly collected among which 123 were tested positive and 102 negatives for SARS-CoV-2 based on RT-PCR. Antigen dosing were performed on a Cobas 8000 e801 analyzer.

resultsThe antigen test diagnosed SARS-CoV-2 infection status with an overall sensitivity of 65,85% (95% CI 56,76-74,16%), a specificity of 100% (95% CI 96,49-100%) with a Cut-off value ≥ 1. When the cut-off value for the antigen assay was set to > 0,673 COI, the accuracy reached its highest level with a sensitivity of 74,8% (95% CI 66,2 - 82,2%) and a specificity of 97,1% (95% CI 91,6 - 99,4%). Imprecision was estimated in accordance with manufacturer's claims.

conclusionsWe obtained an overall sensitivity of 65,85% (95% CI 56,76-74,16%) and a specificity of 100% (95% CI 96,49-100%), slightly higher than the results reported by the manufacturer. Yet, it remains relatively low comparatively to what is generally acceptable for these antigenic assays (a relative sensitivity of 80%). We also noticed that the accuracy could reach its highest level if the cut-off is set above 0,673 which is lower than established by the manufacturer. Thus, our results suggest that the Elecsys® SARS-CoV-2 Antigen assays, should be improved prior to be used in a SARS-Cov-2 screening strategy. However, if one antigenic assay could demonstrate acceptable performance, it might be centralized in clinical laboratories, keeping the RT-PCR in a second phase for confirmation.

Indexed as

COVID-19SARS-CoV-2Antigens, ViralCOVID-19 Serological TestingHumansNasopharynxSensitivity and SpecificityAntigens, ViralAutomated antigen immunoassayCOVID-19SARS-CoV-2 testing strategySensitivity