Evidence map›Paper›PMID 34621186›Full record

ArticleFrontiers in physiology2021

Effects of a Short-Term Left Ventricular Assist Device on Hemodynamics in a Heart Failure Patient-Specific Aorta Model: A CFD Study.

Yu Wang, Junwei Wang, Jing Peng, Mingming Huo, Zhiqiang Yang, Guruprasad A Giridharan, Yong Luan, Kairong Qin

Open access · goldAbstract read
In one paragraph

Article in Frontiers in physiology, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
0.5field-weighted citation impact, top 39% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed, 11 citations in OpenAlex.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 5 institutions in 2 countries.

Yu WangSchool of Optoelectronic Engineering and Instrumentation Science, Dalian University of Technology, Dalian, China.
Junwei WangSchool of Optoelectronic Engineering and Instrumentation Science, Dalian University of Technology, Dalian, China.
Jing PengSchool of Optoelectronic Engineering and Instrumentation Science, Dalian University of Technology, Dalian, China.
Mingming HuoSchool of Optoelectronic Engineering and Instrumentation Science, Dalian University of Technology, Dalian, China.
Zhiqiang YangDepartment of Cardiovascular Computed Tomography (CT) Examination, The First Affiliated Hospital of Dalian Medical University, Dalian, China.
Guruprasad A GiridharanDepartment of Bioengineering, University of Louisville, Louisville, KY, United States.
Yong LuanDepartment of Anesthesiology, The First Affiliated Hospital of Dalian Medical University, Dalian, China.
Kairong QinSchool of Optoelectronic Engineering and Instrumentation Science, Dalian University of Technology, Dalian, China.
Dalian University of Technology · CNDalian University · CNDalian Medical University · CNFirst Affiliated Hospital of Dalian Medical University · CNUniversity of Louisville · US

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Patients with heart failure (HF) or undergoing cardiogenic shock and percutaneous coronary intervention require short-term cardiac support. Short-term cardiac support using a left ventricular assist device (LVAD) alters the pressure and flows of the vasculature by enhancing perfusion and improving the hemodynamic performance for the HF patients. However, due to the position of the inflow and outflow of the LVAD, the local hemodynamics within the aorta is altered with the LVAD support. Specifically, blood velocity, wall shear stress, and pressure difference are altered within the aorta. In this study, computational fluid dynamics (CFD) was used to elucidate the effects of a short-term LVAD for hemodynamic performance in a patient-specific aorta model. The three-dimensional (3D) geometric models of a patient-specific aorta and a short-term LVAD, Impella CP, were created. Velocity, wall shear stress, and pressure difference in the patient-specific aorta model with the Impella CP assistance were calculated and compared with the baseline values of the aorta without Impella CP support. Impella CP support augmented cardiac output, blood velocity, wall shear stress, and pressure difference in the aorta. The proposed CFD study could analyze the quantitative changes in the important hemodynamic parameters while considering the effects of Impella CP, and provide a scientific basis for further predicting and assessing the effects of these hemodynamic signals on the aorta.

Indexed as

CFDheart failurehemodynamic performanceleft ventricular assist devicepatient-specific aorta model

Identifiers

PMID34621186
PMCPMC8491745
OpenAlexW3201176884

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.