ArticleJournal of medical physics
Simultaneous Optimization of Radiation-Imaging Coincidence for a Multi-Energy Linac.
Article in Journal of medical physics. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
No grant is acknowledged in the PubMed record.
Abstract
introductionMedical physics guidelines stress the importance of radiation-imaging coincidence, especially for stereotactic treatments. However, multi-energy linear accelerators may only allow a single imaging isocenter. A procedure was developed to simultaneously optimize radiation-imaging isocenter coincidence for all linac photon energies on a Versa HD. MATERIALS AND
methodsFirst, the radiation beam center of each energy was adjusted to match the collimator rotation axis using a novel method that combined ion chamber measurements with a modified Winston-Lutz (WL) test using images only at gantry, couch, and collimator angles of 0°. With all energies properly steered, an 8-field WL test was performed to determine average linac isocenter position across all energies, gantry, and collimator angles. Lasers and the kV imaging isocenter were calibrated to the average linac isocenter of all photon energies. Finally, A 12-field WL test consisting of gantry, couch, and collimator rotations was used to adjust the couch rotation axis to the average linac isocenter, thereby minimizing overall radiation-imaging isocentricity of the system.
resultsUsing this method, the beam centers were calibrated within 0.10 mm of collimator rotation axis, and linac isocenter coincidence was within 0.20 mm for all energies. Couch isocenter coincidence was adjusted within 0.20 mm of average linac isocenter. Average radiation-imaging isocentricity for all energies was 0.89 mm (0.80-0.98 mm) for a single imaging isocenter.
conclusionThis work provides a method to adjust radiation-imaging coincidence within 1.0 mm for all energies on Elekta's Versa HD.