Evidence map›Paper›PMID 34504935›Full record

ArticleTrauma case reports2021

Technical note on the removal of a "cold-welded" lag screw from a Trigen Meta-Tan nail.

Yohei Yanagisawa, Yusuke Eda, Masashi Yamazaki

Erratum issuedOpen access · goldAbstract readCase Reports
In one paragraph

Article in Trauma case reports, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. An erratum has been issued. Cited by 6 papers.

0numbers the graph read from it
0cells of the map it votes in
6citing papers in PubMed
1.0field-weighted citation impact, top 23% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

6 citing papers in PubMed, 8 citations in OpenAlex.

  1. Review
  2. Article
  3. Article
  4. Article
  5. Article
  6. Article
4 · The record

Corrections and comments

5 · Who and what money

Authors and funding

3 authors at 1 institution in 1 country.

Yohei YanagisawaDepartment of Emergency and Critical Care Medicine, University of Tsukuba Hospital, 2-1-1 Amakubo, Tsukuba, Ibaraki 305-8576, Japan.
Yusuke EdaDepartment of Orthopaedic Surgery, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki 305-8575, Japan.
Masashi YamazakiDepartment of Orthopaedic Surgery, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki 305-8575, Japan.
University of Tsukuba · JP

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Intramedullary or cephalomedullary nail removal often is performed during nonunion reoperations. We have experienced a rare case in which it was difficult to remove the lag screw of the antegrade intramedullary nail, requiring a large amount of force to be applied over a long period. Removal of the lag screw is essential for removal of the nail and subsequent revision surgery. In our case, the lag screw could be removed only by cutting the screw with a carbide drill. For cases in which the nail and lag screw are firmly fixed, surgeons should prepare for the possibility of their separation using a carbide drill. Written informed consent was obtained from the patient for publication of this case report and accompanying images.

Indexed as

Atypical femoral fractureCarbide drillCold-welded screwLag screwSubtrochanteric femoral fracture

Identifiers

PMID34504935
PMCPMC8416944
OpenAlexW3197464600

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked