Evidence map›Paper›PMID 34487244›Full record

ArticleJournal of materials science. Materials in medicine2021

Evaluation of Ti/Al alloy coated with biogenic hydroxyapatite as an implant device in dogs' femur bones.

E M Mahmoud, M Sayed, M Awaad, S T El-Zomor, M Blum, A Killinger, R Gadow, S M Naga

Open access · goldAbstract readEvaluation Study
In one paragraph

Article in Journal of materials science. Materials in medicine, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
0.3field-weighted citation impact, top 46% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed, 6 citations in OpenAlex.

  1. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 2 institutions in 1 country.

E M MahmoudRefractories, Ceramics and building materials Department, National Research Centre, Cairo, Egypt. dr_eman2006@yahoo.com.ORCID http://orcid.org/0000-0001-6717-9533
M SayedRefractories, Ceramics and building materials Department, National Research Centre, Cairo, Egypt.
M AwaadRefractories, Ceramics and building materials Department, National Research Centre, Cairo, Egypt.
S T El-ZomorDepartment of Surgery, Anaesthesiology and Radiology, Faculty of Veterinary Medicine, Cairo University, Cairo, Egypt.
M BlumInstitute for Manufacturing Technologies of Ceramic Components and Composites (IMTCCC), Stuttgart University, Stuttgart, Germany.
A KillingerInstitute for Manufacturing Technologies of Ceramic Components and Composites (IMTCCC), Stuttgart University, Stuttgart, Germany.
R GadowInstitute for Manufacturing Technologies of Ceramic Components and Composites (IMTCCC), Stuttgart University, Stuttgart, Germany.
S M NagaRefractories, Ceramics and building materials Department, National Research Centre, Cairo, Egypt.
National Research Centre · EGCairo University · EG

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

The main target of the present research was a full assessment of the toxicity effects and biocompatibility of a Ti/Al-alloy device coated with biogenic hydroxyapatite (bHA) when implanted in dogs in comparison with those of an uncoated Ti/Al-alloy device. The coating of the alloy was carried out using controlled high-velocity suspension flame spray (HVSFS) technique. Both coated and uncoated devices were implanted in dogs' femur bones for different time periods (45 days and 90 days). Bone-formation ability and healing were followed up, and blood analysis was performed, at Time zero (immediately post surgery), and then at 3 days, 45 days, and 90 days post surgery. Bone mineral density checks, radiological scans of the femur bone, and histological analysis were also conducted. The in-vivo study results proved that implantation of a device made from bHA-coated Ti/Al alloy in dogs' femur bones is completely safe. This is due to the high osteoconductivity of the coated alloy, which enables the formation of new bone and a full connection between new and original bone material. At 90 days post surgery, the coated alloy had been completely digested within the original bone; thus, it appeared as a part of the femur bone and not as a foreign body. Both the scanning electron microscopy with energy-dispersive X-ray and histology analysis findings affirmed the results. Furthermore, the blood tests indicated no toxicity effects during the 90 days of implantation.

Indexed as

Prostheses and ImplantsAlloysAluminumAnimalsBone DensityBone SubstitutesCoated Materials, BiocompatibleDogsDurapatiteFemurMaleMicroscopy, Electron, ScanningOsseointegrationOsteogenesisRandom AllocationSurface PropertiesAlloysAluminumBone SubstitutesCoated Materials, BiocompatibleDurapatiteTitanium

Identifiers

PMID34487244
PMCPMC8421309
OpenAlexW3196827690

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.