ArticleJournal of thoracic disease2021
A comparative study of the RuiBreath and NIOX VERO analyzers for detecting fractional exhaled nitric oxide.
Article in Journal of thoracic disease, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed, 1 citations in OpenAlex.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundFractional exhaled nitric oxide (FeNO) measurement is a reliable, noninvasive marker of airway inflammation. Portable FeNO analyzers facilitate the assessment of airway inflammation in primary care. Differences between analyzers from different manufacturers are not comparable. Here, we aimed to compare the FeNO values obtained by a new portable device (RuiBreath, Guangzhou Ruipu Medical Technology Co., Ltd, Guangzhou, China) to those obtained by the widely used NIOX VERO portable analyzer (Aerocrine AB, Solna, Sweden) in patients with asthma.
methodsThis prospective validation study enrolled patients (≥14 years old) with asthma over a 2-month period (July and August 2019) at the Beijing Chao-Yang Hospital. At least one valid FeNO measurement was obtained using each analyzer for all the participants.
resultsThere were 197 participants in this study. The FeNONIOX and FeNORuiBreath values significantly differed (P=0.016). After log-transformation, a difference was found only when the FeNONIOX was <25 ppb (P<0.001). The FeNONIOX and FeNORuiBreath values had a significant correlation (r=0.938, P<0.001), which was confirmed by the Altman-Bland plot. Using a receiver-operating characteristic curve analysis, when using 49 ppb as the cut-off point for the two devices in identifying patients with symptomatic asthma symptoms, the sensitivity and specificity were 0.42 and 0.88, respectively, by NIOX, and 0.40 and 0.89, respectively, by RuiBreath.
conclusionsThis is the first report of FeNO values obtained by the new portable RuiBreath FeNO analyzer. The FeNORuiBreath values are reliable and directly comparable with the FeNONIOX values.