ArticleMedical physics2021
Development of a storage phosphor imaging system for proton pencil beam spot profile determination.
Article in Medical physics, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
1 citing paper in PubMed.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors.
Funding
Abstract
purposeAccurate two-dimensional (2D) profile measurements at submillimeter precision are necessary for proton beam commissioning and periodic quality assurance (QA) purposes and are currently performed at our institution with a commercial scintillation detector (Lynx PT) with limited means for independent checks. The purpose of this work was to create an independent dosimetry system consisting of an in-house optical scanner and a BaFBrI:Eu
methodsAn in-house 2D laboratory optical scanner was constructed and spatially calibrated for accurate 2D photostimulated luminescence (PSL) dosimetry. Square 5 × 5 cm
resultsThe physical scanner settings of (a) Δτ = 27 ms time interval between data samples, (b) v
conclusionOur combined dosimetric system including an in-house laboratory optical scanner and reusable BaFBr