ArticlePhysics and imaging in radiation oncology2021
Patient position verification in magnetic-resonance imaging only radiotherapy of anal and rectal cancers.
Article in Physics and imaging in radiation oncology, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed, 1 synthesis or guideline pooled it.
- Artificial Intelligence Supported Analysis of Anal Sphincter and Levator Ani Muscle Using Medical Imaging Techniques: A Systematic Review.International urogynecology journal · 2026Pooled it
- Cone beam computed tomography image guidance within a magnetic resonance imaging-only planning workflow.Physics and imaging in radiation oncology · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
14 authors.
Funding
Abstract
background and purposeMagnetic resonance (MR)-only treatment pathways require either the MR-simulation or synthetic-computed tomography (sCT) as an alternative reference image for cone beam computed tomography (CBCT) patient position verification. This study assessed whether using T2 MR or sCT as CBCT reference images introduces systematic registration errors as compared to CT for anal and rectal cancers. MATERIALS AND
methodsA total of 32 patients (18 rectum,14 anus) received pre-treatment CT- and T2 MR- simulation. Routine treatment CBCTs were acquired. sCTs were generated using a validated research model. The local clinical registration protocol, using a grey-scale registration algorithm, was performed for 216 CBCTs using CT, MR and sCT as the reference image. Linear mixed effects modelling identified systematic differences between modalities.
resultsSystematic translation and rotation differences to CT for MR were -0.3 to + 0.3 mm and -0.1 to 0.4° for anal cancers and -0.4 to 0.0 mm and 0.0 to 0.1° for rectal cancers, and for sCT were -0.4 to + 0.8 mm, -0.1 to 0.2° for anal cancers and -0.6 to + 0.2 mm, -0.1 to + 0.1° for rectal cancers.
conclusionsT2 MR or sCT can successfully be used as reference images for anal and rectal cancer CBCT position verification with systematic differences to CT <±1 mm and <±0.5°. Clinical enabling of alternative modalities as reference images by vendors is required to reduce challenges associated with their use.