Droplet Size and Distribution of Nebulized 3% Sodium Chloride, Albuterol, and Epoprostenol by Phase Doppler Particle Analyzer.

Kelly McDermott, Jason G Oakley

Abstract read
In one paragraph

Article in Current therapeutic research, clinical and experimental, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed.

  1. Article
  2. Tracheomalacia Reduces Aerosolized Drug Delivery to the Lung.Journal of aerosol medicine and pulmonary drug delivery · 2024
    Article
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

2 authors.

Kelly McDermottUniversity of Wisconsin Hospitals and Clinics, Madison, Wisconsin.
Jason G OakleyDepartment of Mechanical Engineering, University of Wisconsin-Madison College of Engineering, Madison, Wisconsin.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundAerosolized drug therapy administered to mechanically ventilated patients is a standard part of pulmonary critical care medicine. Aerosol particle size and distribution are important factors in the optimal delivery of aerosolized drugs to ventilated patients.

objectiveThe objective of this study was to characterize aerosol droplet size and distribution with laser diffraction for nebulized 3% sodium chloride, albuterol, and epoprostenol sodium (containing glycine) delivered via Aeroneb Solo Mesh Nebulizers (Aerogen, Mountain View, California).

methodsA series of functional flow tests were run on each of 8 Solo mesh nebulizers before the study to verify accuracy of flow rates in milliliters per minute. Aerosolized droplets exiting the nebulizer heads were then measured using a phase Doppler particle analyzer. Data collected during delivery of 3% sodium chloride, albuterol, and epoprostenol sodium included aerosol droplet size distribution, mass median aerodynamic diameter (MMAD), and geometric standard deviation. For each Solo nebulizer, droplet size measurements were taken 2 cm away from the nebulizer head and 2 cm away from the wye of a heated, humidified adult ventilator circuit. For measurements taken at the wye, 4 distinct, continuous flow rates (2, 10, 20, and 40 L/min) were generated by an air pump to simulate inspiratory flows delivered with mechanical ventilation. The expiratory limb was capped, and the nebulizer head was inserted into the breathing circuit upstream of the humidifier.

resultsEach Solo nebulizer met Aerogen's recommended minimum flow rate of 0.2 mL/min, ranging from 0.23 to 0.31 mL/min. The MMAD of the 3 tested aerosols was several times smaller when measured at the wye outlet of the heated/humidified breathing circuit (0.82-2.73 µm) compared with droplets measured directly at the nebulizer outlet (MMAD, 4.6-7.3 µm). There was also significant variability across Solo heads with some ventilator flow rates. The mean MMAD at the wye for the 3% sodium chloride solution, albuterol, and epoprostenol test solutions was 1.62 µm, 1.09 µm, and 1.18 µm, respectively. The mean MMAD at the nebulizer for the 3% sodium chloride solution, albuterol, and epoprostenol test solutions was 5.37 µm, 5.73 µm, and 6.73 µm, respectively.

conclusionsResults from this study suggest that particle size of aerosolized drugs administered via a commonly used setup for delivery of in-line aerosols to mechanically ventilated patients may be several times smaller than expected and may result in less drug being delivered to the patient than previously realized.(