A T1-based correction method for proton resonance frequency shift thermometry in breast tissue.
Article in Medical physics, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed.
- Evaluating the accuracy of the variable flip angle technique for temperature measurement near metallic implants.Quantitative imaging in medicine and surgery · 2025Article
- In vivo simultaneous proton resonance frequency shift thermometry and single reference variable flip angle TMagnetic resonance in medicine · 2025Article
- Magnetic resonance imaging-guided radiofrequency ablation of breast cancer: a current state of the art review.Diagnostic and interventional radiology (Ankara, Turkey) · 2024Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors.
Funding
Abstract
purposeDevelop and evaluate the effectiveness of a T1-based correction method for errors in proton resonant frequency shift thermometry due to non-local field effects caused by heating in fatty breast tissues.
methodsComputational models of human breast tissue were created by segmenting MRI data from a healthy human volunteer. MR-guided focused ultrasound (MRgFUS) heating and MR thermometry measurements were simulated in several locations in the heterogeneous segmented breast models. A T1-based correction method for PRF thermometry errors was applied and the maximum positive and negative errors and the root mean squared error (RMSE) in a region around each heating location was evaluated with and without correction. The method uses T1 measurements to estimate the temperature change in fatty tissues and correct for their influence. Experimental data from a heating study in cadaver breast tissue were analyzed, and the expected PRFS error computed.
resultsThe simulated MR thermometry had maximum single voxel errors ranging between 10% and 18% when no correction was applied. Applying the correction led to a considerable improvement, lowering the maximum error range to 2%-5%. The 5th to 95th percentile interval of the temperature error distribution was also lowered with correction, from approximately 3.5 to 1°C. This correction worked even when T1 times were uniformly raised or lowered by 5%-10%. The experimental data showed predicted errors of 15%.
conclusionsThis simulation study demonstrates that the T1-based correction method reduces MR thermometry errors due to non-local effects from heating in fatty tissues, potentially improving the accuracy of thermometry measurements during MRgFUS treatments. The presented correction method is reliant on having a patient-specific 3D model of the breast, and may be limited by the accuracy of the fat temperatures which in turn may be limited by noise or bias present in the T1 measurements.