ArticleInternational journal of laboratory hematology2021
A performance evaluation of chemiluminescence enzyme immunoassays on the Sysmex CN-6500 haemostasis analyser.
Article in International journal of laboratory hematology, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
6 citing papers in PubMed, 5 citations in OpenAlex.
- Molecular biomarkers in dialysis-related thrombosis and anticoagulation monitoring: thrombin-antithrombin complex, plasmin-α2-antiplasmin complex, soluble thrombomodulin, tPA/PAI-1/tPAIC, and Anti-Xa.Frontiers in medicine · 2026Review
- Comparison of SYSMEX CN-6000 and STAGO STA-R: Effects of hemolysis and lipemia.Journal of medical biochemistry · 2025Article
- Expression of Recombinant Tissue Plasminogen Activator - Plasminogen Activator Inhibitor Complex in Mammal Cells as Quality Control Materials in Immunoassay.The protein journal · 2025Article
- A clinical study on the screening efficacy of six thrombotic markers TAT, PIC, TM, t-PAI-C, D-D, and FDP for hypercoagulable state in pregnant women.Pakistan journal of medical sciences · 2025Article
- Automated measurement of coagulation and fibrinolytic activation markers: Outcomes in coronavirus disease 2019 (COVID-19) patients.International journal of laboratory hematology · 2022Observational
- A performance evaluation of chemiluminescence enzyme immunoassays on the Sysmex CN-6500 haemostasis analyser.International journal of laboratory hematology · 2021Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundThe Sysmex CN-6500 is a new haemostasis analyser with an integrated immunoassay module that performs chemiluminescence enzyme assay (CLEIA) in addition to coagulation, turbidimetric, chromogenic and platelet aggregation tests.
aimsTo evaluate the analytical performance of the CN-6500 against the predicate device (Sysmex HISCL-800) for soluble thrombomodulin (TM), thrombin-antithrombin (TAT), tissue plasminogen activator/plasminogen activator inhibitor 1 complex (tPAI-C) and plasmin α2 plasmin inhibitor complex (PIC) assays.
methodsImprecision was assessed by testing two levels of quality control plasmas 10 times on 5 separate days. Comparability was studied in 230 plasmas from normal donors (n = 30), patients with suspected disseminated intravascular coagulation (DIC, n = 100), sepsis (n = 20) or liver disease (n = 20), lipaemic (n = 20), haemolysed (n = 20) and icteric samples (n = 20). Limit of detection, limit of quantitation and linearity were determined by testing serial dilutions of normal plasma. Sample carryover was assessed by testing samples with high and low normal levels of the analytes concerned.
resultsThe CN-6500 performed 21 CLEIA tests per hour, while simultaneously performing coagulation tests. Acceptable between-run imprecision was obtained using commercial controls with normal and high activity for each analyte (%CV <4%), for all four assays. Excellent linearity was observed (slope 0.89-1.03; r
conclusionsOur data demonstrate that the performance of the CLEIA assays on the CN-6500 is comparable to that of a stand-alone immunoassay analyser.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials
Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.