ArticleMedical physics2021

Motion artifacts assessment and correction using optical tracking in synchrotron radiation breast CT.

Luca Brombal, Lucia Mariel Arana Peña, Fulvia Arfelli, Renata Longo, Francesco Brun, Adriano Contillo, Francesca Di Lillo, Giuliana Tromba, Vittorio Di Trapani, Sandro Donato and 2 more

Abstract read
In one paragraph

Article in Medical physics, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

12 authors.

Luca BrombalDepartment of Physics, University of Trieste, Trieste, Italy.
Lucia Mariel Arana PeñaDepartment of Physics, University of Trieste, Trieste, Italy.
Fulvia ArfelliDepartment of Physics, University of Trieste, Trieste, Italy.
Renata LongoDepartment of Physics, University of Trieste, Trieste, Italy.
Francesco BrunDivision of Trieste, Istituto Nazionale di Fisica Nucleare, Trieste, Italy.
Adriano ContilloElettra-Sincrotrone Trieste S.C.p.A., Trieste, Italy.
Francesca Di LilloElettra-Sincrotrone Trieste S.C.p.A., Trieste, Italy.
Giuliana TrombaElettra-Sincrotrone Trieste S.C.p.A., Trieste, Italy.
Vittorio Di TrapaniDepartment of Physical sciences, Earth and environment, University of Siena, Siena, Italy.
Sandro DonatoDepartment of Physics, University of Calabria, Arcavacata di Rende, Cosenza, Italy.
Ralf Hendrik MenkDivision of Trieste, Istituto Nazionale di Fisica Nucleare, Trieste, Italy.
Luigi RigonDepartment of Physics, University of Trieste, Trieste, Italy.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeThe SYRMA-3D collaboration is setting up a breast computed tomography (bCT) clinical program at the Elettra synchrotron radiation facility in Trieste, Italy. Unlike the few dedicated scanners available at hospitals, synchrotron radiation bCT requires the patient's rotation, which in turn implies a long scan duration (from tens of seconds to few minutes). At the same time, it allows the achievement of high spatial resolution. These features make synchrotron radiation bCT prone to motion artifacts. This article aims at assessing and compensating for motion artifacts through an optical tracking approach.

methodsIn this study, patients' movements due to breathing have been first assessed on seven volunteers and then simulated during the CT scans of a breast phantom and a surgical specimen, by adding a periodic oscillatory motion (constant speed, 1 mm amplitude, 12 cycles/minute). CT scans were carried out at 28 keV with a mean glandular dose of 5 mGy. Motion artifacts were evaluated and a correction algorithm based on the optical tracking of fiducial marks was introduced. A quantitative analysis based on the structural similarity (SSIM) index and the normalized mean square error (nMSE) was performed on the reconstructed CT images.

resultsCT images reconstructed through the optical tracking procedure were found to be as good as the motionless reference image. Moreover, the analysis of SSIM and nMSE demonstrated that an uncorrected motion of the order of the system's point spread function (around 0.1 mm in the present case) can be tolerated.

conclusionsResults suggest that a motion correction procedure based on an optical tracking system would be beneficial in synchrotron radiation bCT.

Indexed as

ArtifactsSynchrotronsAlgorithmsBreastHumansPhantoms, ImagingTomography, X-Ray Computedbreast CTmotion artifactsoptical trackingphase contrast

Identifiers

PMID34252212
PMCPMC9291820

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.