Evidence map›Paper›PMID 34196991›Full record

Trial reportJournal of cardiovascular electrophysiology2021

Ultralow temperature cryoablation using near-critical nitrogen for cavotricuspid isthmus-ablation, first-in-human results.

Martijn N Klaver, Tom J R De Potter, Konstantinos Iliodromitis, Alexander Babkin, David Cabrita, Davide Fabbricatore, Lucas V A Boersma

2 registry-linked trialsOpen access · hybridAbstract readClinical TrialMulticenter Study
In one paragraph

Trial report in Journal of cardiovascular electrophysiology, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. It is linked to 2 registered trials, which are not on this map. Cited by 8 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
8citing papers in PubMed, 1 pooled it
2.7field-weighted citation impact, top 9% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

NCT02355106 nacompletednot on this map

A Prospective, Multicenter, Investigation of the Adagio PolarStar System in Subjects With Atrial Flutter

TypeinterventionalSponsorAdagio MedicalRan2015 to 2019Enrolled17ConditionsAtrial FlutterArmsPolarStart System
NCT02839304 nacompletednot on this map

A Prospective, Multicenter Investigation of the Adagio Cryoablation System in Patients With Atrial Fibrillation

TypeinterventionalSponsorAdagio MedicalRan2016 to 2021Enrolled80ConditionsAtrial Fibrillation, Atrial FlutterArmsCryoablation System
3 · Its place in the literature

Who cites it

8 citing papers in PubMed, 1 synthesis or guideline pooled it, 17 citations in OpenAlex.

  1. Esophageal injury, perforation, and fistula formation following atrial fibrillation ablation.Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2024
    Pooled it
  2. Trial
  3. Article
  4. Review
  5. Article
  6. Article
  7. Article
  8. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 3 institutions in 3 countries.

Martijn N KlaverDepartment of Cardiology, St. Antonius Hospital, Nieuwegein, The Netherlands.ORCID 0000-0002-0215-6637
Tom J R De PotterDepartment of Cardiology, Onze-Lieve-Vrouw Ziekenhuis, Aalst, Belgium.ORCID 0000-0003-1424-114X
Konstantinos IliodromitisDepartment of Cardiology, Onze-Lieve-Vrouw Ziekenhuis, Aalst, Belgium.ORCID 0000-0003-4243-9823
Alexander BabkinAdagio Medical, Laguna Hills, California, USA.
David CabritaAdagio Medical, Laguna Hills, California, USA.
Davide FabbricatoreDepartment of Cardiology, Onze-Lieve-Vrouw Ziekenhuis, Aalst, Belgium.ORCID 0000-0002-5622-3233
Lucas V A BoersmaDepartment of Cardiology, St. Antonius Hospital, Nieuwegein, The Netherlands.ORCID 0000-0003-4591-4220
Onze Lieve Vrouwziekenhuis Hospital · BEAdagio Health · USSt. Antonius Ziekenhuis · NL

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

introductionCryoablation has evolved as a safe alternative to radiofrequency ablation in the treatment of several supraventricular arrhythmias and has potential advantages, yet is limited by the properties of the cryogen used. We investigated a novel ultralow temperature cryoablation (ULTC) system using nitrogen near its liquid-vapor critical point as a freezing source, achieving temperatures as low as -196 degrees Celsius in a long linear catheter with a continuous energy release. Initial safety, procedural and efficacy outcomes of ULTC are described in patients undergoing cavotricuspid isthmus (CTI) ablation. METHODS AND

resultsThe Cryocure studies (NCT02355106, NCT02839304) are prospective, single-arm, multi-center, first-in-human clinical studies in 17 patients with atrial flutter (AFL) and 13 patients with atrial fibrillation (AF). A total of 30 patients, mean age 65 ± 8 years old and 67% male, were enrolled and underwent ablation of the CTI. Acute success, defined as the confirmation of stable bidirectional conduction block across the CTI, was achieved in all 30 patients. After 12 months of follow-up, 14 out of 17 AFL patients remained free from any AFL. One (3.3%) procedure-related but not device-related serious adverse event was reported, involving transient inferolateral ST-elevation associated with temporary AV conduction block.

conclusionIn this first-in-human clinical study the safety and performance results demonstrate the capabilities of ultralow temperature near-critical nitrogen as an effective energy source for CTI ablation. Ongoing, larger, studies should confirm our findings and evaluate the capabilities to create linear and focal transmural lesions in other arrhythmias.

Indexed as

Atrial FlutterCatheter AblationCryosurgeryAgedFemaleHumansMaleMiddle AgedNitrogenProspective StudiesTemperatureTreatment OutcomeNitrogencavotricuspid isthmuscryoablationfirst-in-humannear-critical nitrogenultralow temperature

Identifiers

PMID34196991
PMCPMC8457224
OpenAlexW3174014749

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials