Evidence map›Paper›PMID 34178173›Full record

ReviewBiophysical reviews2021

Electrodes in external electrohysterography: a systematic literature review.

Thierry R Jossou, Aziz Et-Tahir, Zakaria Tahori, Abdelmajid El Ouadi, Daton Medenou, Abdelmajid Bybi, Latif Fagbemi, Mohamed Sbihi, Davide Piaggio

Open access · greenAbstract readReview
In one paragraph

Review in Biophysical reviews, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
0.6field-weighted citation impact, top 38% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 11 citations in OpenAlex.

  1. Article
  2. Biophysical reviews · 2021
    Article
  3. Biophysical reviews · 2021
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors at 4 institutions in 3 countries.

Thierry R JossouDepartment of Biomedical Engineering, Ecole Polytechnique d'Abomey-Calavi, University of Abomey-Calavi, Abomey-Calavi, Benin.ORCID 0000-0002-7229-5122
Aziz Et-TahirMaterials, Energy, Acoustics Team, Ecole Supérieure de Technologie de Salé, University Mohammed V, Rabat, Morocco.
Zakaria TahoriENSAK, Universite Ibn Tofail Kenitra, Kenitra, Morocco.
Abdelmajid El OuadiENSAK, Universite Ibn Tofail Kenitra, Kenitra, Morocco.
Daton MedenouDepartment of Biomedical Engineering, Ecole Polytechnique d'Abomey-Calavi, University of Abomey-Calavi, Abomey-Calavi, Benin.
Abdelmajid BybiMaterials, Energy, Acoustics Team, Ecole Supérieure de Technologie de Salé, University Mohammed V, Rabat, Morocco.
Latif FagbemiDepartment of Biomedical Engineering, Ecole Polytechnique d'Abomey-Calavi, University of Abomey-Calavi, Abomey-Calavi, Benin.
Mohamed SbihiMaterials, Energy, Acoustics Team, Ecole Supérieure de Technologie de Salé, University Mohammed V, Rabat, Morocco.
Davide PiaggioSchool of Engineering, University of Warwick, Coventry, CV4 7AL UK.
Mohammed V University · MAUniversité d'Abomey-Calavi · BJUniversité Ibn-Tofail · MAUniversity of Warwick · GB

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundIn low-income countries, pregnant women do not have easy access to health care, especially in rural and peri-urban areas. In this context, they can be surprised by the uterine contractions that precede childbirth and sometimes find themselves giving birth at home or on the way to the nearest health facility (located miles away from their home). In view of the development of an external uterine electrohysterogram acquisition system for labour prediction, a review of the literature on electrodes and their characteristics is necessary.

methodsA comprehensive literature review was conducted to collate information on the use of electrodes in external EHG recording and their characteristics.

resultsWet electrodes based on Ag/AgCl redox chemistry are the most common type of electrodes for EHG, employed in different configurations on the pregnant woman's abdomen. All positioning configurations are around the vertical median axis if they are not placed directly on it. Positioning below the navel seems to be the most efficient. The number of source, reference, and ground electrodes used varies from one author to another, as does the distance between the electrodes.

conclusionTwo well-positioned source electrodes on the vertical median axis, with ground electrode on the right side of the hip and reference one on the left side, are able to generate a good external EHG recording signal. The minimum allowed inter-electrode distance is approximately 17.5 to 25mm.

Indexed as

DeliveryElectrodesElectrohysterographyLabourRecording

Identifiers

PMID34178173
PMCPMC8214640
OpenAlexW3162530402

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.