ArticleJournal of clinical laboratory analysis2021
Interference of eltrombopag with bilirubin measurements on the Vitros 5600 analyzer.
Article in Journal of clinical laboratory analysis, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
3 citing papers in PubMed, 4 citations in OpenAlex.
- Analytical pitfalls in dry and wet chemistry platforms: From a trainee perspective.Practical laboratory medicine · 2026Article
- Thrombopoietin Receptor Agonists for Thrombocytopenia in Pediatric Hematologic Malignancies.Pediatric blood & cancer · 2025Article
- Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
4 authors at 2 institutions in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundEltrombopag is a thrombopoietin-receptor agonist used to restore platelet count to hemostatic levels in chronic immune thrombocytopenia. The drug has shown to have hepatobiliary adverse effects, but also positive interference with the analytical measurement of bilirubin. Understanding the degree of interference of this drug with bilirubin testing becomes relevant in the clinical management of these patients.
methodsEltrombopag at concentrations ranging from 10 to 150 µg/ml was spiked into plasma samples with different baseline concentrations of bilirubin. Total bilirubin, conjugated, and unconjugated bilirubin were measured for each sample using VITROS TBILI and BuBc slides on the Vitros 5600 automated chemistry platform, and interference was assessed.
resultsPlasma samples spiked with eltrombopag yielded falsely elevated bilirubin measurements compared to baseline, with the degree of elevation increasing with greater concentrations of eltrombopag. Bilirubin values were increased relative to baseline across all groups, except in conjugated bilirubin measurements in samples with low baseline conjugated bilirubin. For samples with low total bilirubin at baseline, >100 µg/ml of eltrombopag resulted in an error of >+0.6 mg/dl on the measured total bilirubin. For samples with low unconjugated bilirubin at baseline, the error for the same concentrations was >+0.7 mg/dl.
conclusionOur results show that, at supra-physiologically high concentrations, eltrombopag can positively interfere with bilirubin measurements on Vitros 5600 platform.