ArticleBiomedical spectroscopy and imaging2020
The Potential of Utilizing Mid-Energy X-Rays for In-Line Phase Sensitive Breast Cancer Imaging.
Article in Biomedical spectroscopy and imaging, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
0 citing papers in PubMed, 4 citations in OpenAlex.
No citing paper in PubMed yet.
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
9 authors at 3 institutions in 1 country.
Funding
Abstract
objectiveThe objective of this study is to demonstrate the potential of utilizing mid-energy x-rays for in-line phase-sensitive breast cancer imaging by phantom studies.
methodsThe midenergy (50-80kV) in-line phase sensitive imaging prototype was used to acquire images of the contrast-detail mammography (CDMAM) phantom, an ACR accreditation phantom, and an acrylic edge phantom. The low-dose mid-energy phase-sensitive images were acquired at 60 kV with a radiation dose of 0.9 mGy, while the high-energy phase-sensitive images were acquired at 90 kV with a radiation dose of 1.2 mGy. The Phase-Attenuation Duality (PAD) principle for soft tissue was used for the phase retrieval. A blind observer study was conducted and paired-sample T-test were performed to compare the mean differences in the two imaging systems.
resultsThe correct detection ratio for the CDMAM phantom for phase-contrast images acquired by the low-dose mid-energy system was 56.91%, whereas images acquired by the high-energy system correctly revealed only 40.97% of discs. The correct detection ratios were 57.88% and 43.41% for phase-retrieved images acquired by the low-dose mid-energy and high-energy imaging systems, respectively. The reading scores for all three groups of objects in the ACR phantom were higher for the mid energy imaging system as compared to the high-energy system for both phase-contrast and phase- retrieved images. The calculated edge enhancement index (EEI) from the acrylic edge phantom image for the mid-energy system was higher than that calculated for the high-energy imaging system. The quantitative analyses showed a higher Contrast to Noise Ratio (CNR) as well as a higher Figure of Merit (FOM) in images acquired by the low-dose mid-energy imaging system.
conclusionThe PAD based retrieval method can be applied in mid-energy system without remarkably affecting the image quality, and in fact, it improves the lesion detectability with a patient dose saving of 25%.