Evidence map›Paper›PMID 34137316›Full record

ArticleProceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine2021

Endurance testing and finite element simulation of a modified hip stem for integration of an energy harvesting system.

Hans-E Lange, Rainer Bader, Daniel Kluess

Open access · hybridAbstract read
In one paragraph

Article in Proceedings of the Institution of Mechanical Engineers. Part H, Journal of engineering in medicine, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
1.7field-weighted citation impact, top 16% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed, 10 citations in OpenAlex.

  1. Trial
  2. Review
  3. The application of impantable sensors in the musculoskeletal system: a review.Frontiers in bioengineering and biotechnology · 2024
    Review
  4. Article
  5. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors at 1 institution in 1 country.

Hans-E LangeDepartment of Orthopaedics, Rostock University Medical Center, Rostock, Germany.ORCID https://orcid.org/0000-0003-4926-769X
Rainer BaderDepartment of Orthopaedics, Rostock University Medical Center, Rostock, Germany.
Daniel KluessDepartment of Orthopaedics, Rostock University Medical Center, Rostock, Germany.
University of Rostock · DE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Instrumented implants are a promising approach to further improve the clinical outcome of total hip arthroplasties. For the integrated sensors or active functions, an electrical power supply is required. Energy harvesting concepts can provide autonomous power with unlimited lifetime and are independent from external equipment. However, those systems occupy space within the mechanically loaded total hip replacement and can decrease the life span due to fatigue failure in the altered implant. We previously presented a piezoelectric energy harvesting system for an energy-autonomous instrumented total hip stem that notably changes the original implant geometry. The aim of this study was to investigate the remaining structural fatigue failure strength of the metallic femoral implant component in a worst-case scenario. Therefore, the modified hip stem was tested under load conditions based on ISO 7206-4:2010. The required five million cycles were completed twice by all samples (n = 3). Additionally applied cycles with incrementally increased load levels up to 4.7 kN did not induce implant failure. In total, 18 million cycles were endured, outperforming the requirements of the ISO standard. Supplementary finite element analysis was conducted to determine stress distribution within the implant. A high stress concentration was found in the region of modification. The stress level showed an increase compared to the previously evaluated physiological loading situation and was close to the fatigue data from the literature. The stress concentration factor compared to the original geometry amounted to 2.56. The assessed stress level in accordance with the experimental fatigue testing can serve as a maximum reference value for further implant design modifications and optimisations.

Indexed as

Arthroplasty, Replacement, HipHip ProsthesisComputer SimulationFemurFinite Element AnalysisProstheses and ImplantsStress, Mechanicalenergy harvestingfatiguefinite element analysistestingTotal hip replacement

Identifiers

PMID34137316
PMCPMC8649416
OpenAlexW3166378382

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.