Evidence map›Paper›PMID 34136867›Full record

ArticleJSES international2021

New precontoured long locking plate for proximal metadiaphyseal fractures of the humerus: a cadaveric study for its use with the minimally invasive technique.

Max Ekdahl, Cristián Dominguez, Miguel Pinedo, Sebastián López, Vicente Gutiérrez

Abstract read
In one paragraph

Article in JSES international, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed.

  1. 45° helical plates are a valid alternative to straight plates for treatment of proximal humeral shaft fractures.Journal of orthopaedic research : official publication of the Orthopaedic Research Society · 2025
    Article
  2. Article
  3. Article
  4. [Effectiveness comparison of two minimally invasive plate osteosynthesis techniques for proximal humeral shaft fractures].Zhongguo xiu fu chong jian wai ke za zhi = Zhongguo xiufu chongjian waike zazhi = Chinese journal of reparative and reconstructive surgery · 2023
    Article
  5. Anatomical analysis of different helical plate designs for proximal humeral shaft fracture fixation.European journal of trauma and emergency surgery : official publication of the European Trauma Society · 2023
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

5 authors.

Max EkdahlOrthopaedic Surgery Department, Clinica Las Condes, Santiago, Chile.
Cristián DominguezOrthopaedic Surgery Department, Clinica Las Condes, Santiago, Chile.
Miguel PinedoOrthopaedic Surgery Department, Clinica Las Condes, Santiago, Chile.
Sebastián LópezOrthopaedic Surgery Department, Clinica Las Condes, Santiago, Chile.
Vicente GutiérrezOrthopaedic Surgery Department, Clinica Las Condes, Santiago, Chile.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundThe purpose of this study was to identify nerves at risk when using a minimally invasive plate osteosynthesis precontoured long proximal humerus locking plate and to evaluate the risk of injury to deltoid insertion and brachialis muscle.

methodsTen cadaveric upper limb specimens were used. A transdeltoid anterolateral approach was performed proximally and a second anterior approach was performed distally. A 14-hole "low" long precountored ALPS locking plate (Biomet Trauma; Zimmer Biomet, Warsaw, IN, USA) was used. Subsequently, anatomic dissection to measure the anatomic relationship of the plate with the deltoid insertion, with the brachialis muscle, and with the axillary, radial, and musculocutaneous nerves was performed.

resultsThe mean humeral length was 302 mm (standard deviation 52.3, 99% confidence interval: 259.3-344.6). In 6 specimens, the axillary nerve was located at the level of the third row of holes of the plate; in 3 specimens, at the level of the fourth row; and in one specimen, at the level of the second row. The distance between the plate and the musculocutaneous nerve was on average 10.2 mm (standard deviation 4, 99% confidence interval: 6.9-13.5) and between the plate and the radial nerve was on average 7.9 mm (standard deviation 4.7, 99% confidence interval: 4-11.8). The plate pierced the anterior distal fibers of the deltoid in all specimens. In 8 specimens, no brachialis muscle fibers were located under the plate.

conclusionsThe use of the long precontoured 14-hole ALPS locking plate with the minimally invasive plate osteosynthesis technique, previously identifying the axillary and musculocutaneous nerves, is feasible; however, the distances between the plate and the nerves remain low, so caution should be maintained. Despite the curved design of the plate, the deltoid insertion is partially compromised in all cases.

Indexed as

Cadaveric studyMIPOProximal metadiaphyseal fractures

Identifiers

PMID34136867
PMCPMC8178608

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.