Evidence map›Paper›PMID 34116614›Full record

Observational studyJournal of diabetes science and technology2022

Accuracy and Stability of a Subcutaneous Flash Glucose Monitoring System in Critically Ill Patients.

Hiromu Naraba, Tadahiro Goto, Mitsuhiro Tokuda, Tomohiro Sonoo, Hidehiko Nakano, Yuji Takahashi, Hideki Hashimoto, Kensuke Nakamura

Open access · bronzeAbstract readObservational Study
In one paragraph

Observational study in Journal of diabetes science and technology, 2022. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
0.4field-weighted citation impact, top 36% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 5 citations in OpenAlex.

  1. Article
  2. Review
  3. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 2 institutions in 1 country.

Hiromu NarabaDepartment of Emergency and Critical Care Medicine, Hitachi General Hospital, Hitachi, Ibaraki, Japan.ORCID 0000-0002-5663-2213
Tadahiro GotoTXP Medical Co., Ltd., University of Tokyo, Bunkyo, Tokyo, Japan.
Mitsuhiro TokudaDepartment of Emergency and Critical Care Medicine, Hitachi General Hospital, Hitachi, Ibaraki, Japan.
Tomohiro SonooDepartment of Emergency and Critical Care Medicine, Hitachi General Hospital, Hitachi, Ibaraki, Japan.
Hidehiko NakanoDepartment of Emergency and Critical Care Medicine, Hitachi General Hospital, Hitachi, Ibaraki, Japan.
Yuji TakahashiDepartment of Emergency and Critical Care Medicine, Hitachi General Hospital, Hitachi, Ibaraki, Japan.
Hideki HashimotoDepartment of Emergency and Critical Care Medicine, Hitachi General Hospital, Hitachi, Ibaraki, Japan.
Kensuke NakamuraDepartment of Emergency and Critical Care Medicine, Hitachi General Hospital, Hitachi, Ibaraki, Japan.
Hitachi General Hospital · JPTokyo Medical University · JP

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundFlash glucose monitoring (FGM) systems can reduce glycemic variability and facilitate blood glucose management within the target range. However, in critically ill patients, only small (

methodThis was a single-center, retrospective observational study. We included a total of 116 critically ill patients who underwent FGM for glycemic control. The accuracy of FGM was assessed as follows using blood gas glucose values as a reference: (1) numerical accuracy using the mean absolute relative difference, (2) clinical accuracy using consensus error grid analysis, and (3) stability of accuracy assessing 14-day trends in consensus error grid distribution.

resultsFGM sensors remained in situ for a median of 6 [4, 11] days. We compared 2014 pairs of measurements between the sensor and blood gas analysis. Glucose values from the sensor were consistently lower, with a mean absolute relative difference of 13.8% (±16.0%), than those from blood gas analysis. Consensus error grid analysis demonstrated 99.4% of the readings to be in a clinically acceptable accuracy zone. The accuracy of FGM was stable across the 14 days after device insertion.

conclusionsFGM had acceptable reliability and accuracy to arterial blood gas analysis in critically ill patients. In addition, the accuracy of FGM persisted for at least 14 days. Our study promotes the potential usefulness of FGM for glycemic monitoring in critically ill patients.

Indexed as

Blood GlucoseBlood Glucose Self-MonitoringCritical IllnessGlucoseHumansReproducibility of ResultsBlood GlucoseGlucosecritical care medicineflash glucose monitoringglycemic controlmean absolute relative difference

Identifiers

PMID34116614
PMCPMC9445337
OpenAlexW3168950293

What OpenQuestion holds

Textmetadata
Read underepoch 390

Registered trials

None linked