ArticleHeart rhythm O22020
Ineffective implantable cardioverter-defibrillator shocks among patients on continuous left ventricular assist device support: Clinical characteristics and management.
Article in Heart rhythm O2, 2020. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed, 5 citations in OpenAlex.
- Body Piercing with a Metallic Tongue Stud Resulting in Ineffective Implantable Cardioverter-defibrillator Shocks: "Heart to Mouth".The Journal of innovations in cardiac rhythm management · 2021Article
- Azygous Vein Coil Implantation in Left Ventricular Assist Device Patients: A Hands-on Approach.The Journal of innovations in cardiac rhythm management · 2021Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
6 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
backgroundPlacement of a left ventricular assist device (LVAD) has been described to compromise implantable cardioverter-defibrillator (ICD) defibrillation threshold (DFT). Elevated DFT will have negative consequences and increases the risk of ineffective ICD shocks, morbidity, and mortality. DFT testing is not routinely performed in clinical practice, despite this fact.
objectiveWe describe the clinical characteristics of 7 LVAD patients who presented with multiple ineffective ICD shocks, along with the management strategy in such patients.
methodsSeven patients (5 male, mean age 52.2 ± 9 years, 85.7% nonischemic cardiomyopathy) with an ICD in situ who progressed to NYHA class IV, ACC/AHA stage D chronic systolic congestive heart failure who underwent successful LVAD implantation presented to our institution in the setting of ventricular tachyarrhythmia and ineffective ICD shocks. Six patients underwent implantation of azygos and subclavian coils with subsequent DFT testing. The remaining patient was made comfort care.
resultsFive patients had successful DFT testing with azygos (n = 4) and subclavian (n = 1) defibrillation coil implantation. One patient had unsuccessful DFT testing despite evaluation of multiple shock vectors. There were no major or minor vascular complications in any of the cases. There were no procedural-related deaths.
conclusionThis case series highlights the need for a systematic approach to management of ICDs and DFT testing in LVAD patients. The addition of new shock vectors with azygos and subclavian coil implantation may reduce DFT, shock burden, morbidity, and mortality.
Indexed as
Identifiers
What OpenQuestion holds
Registered trials