Operation and calibration of the novel PTW 1600SRS detector for the verification of single isocenter stereotactic radiosurgery treatments of multiple small brain metastases.
Article in The British journal of radiology, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 6 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
6 citing papers in PubMed, 14 citations in OpenAlex.
- Deep learning-based upsampling of 2D detector array measurements for patient plan verification in radiotherapy.Medical physics · 2026Article
- Evaluation of 2D ion chamber arrays for patient specific quality assurance using a static phantom at a 0.35 T MR-Linac.Zeitschrift fur medizinische Physik · 2025Article
- Accuracy of Beamscan Software in Determining the Inflection Point from the FFF Beam Profile Using Several Array Detectors.Asian Pacific journal of cancer prevention : APJCP · 2024Article
- Commissioning and clinical evaluation of a novel high-resolution quality assurance digital detector array for SRS and SBRT.Journal of applied clinical medical physics · 2024Article
- Comprehensive characterisation of the IBA myQA SRS for SRS and SBRT patient specific quality assurance.Physical and engineering sciences in medicine · 2024Article
- Evaluation of a high resolution diode array for CyberKnife quality assurance.Journal of applied clinical medical physics · 2023Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 1 institution in 1 country.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectivesThe aim of this work was to evaluate the operation of the 1600SRS detector and to develop a calibration procedure for verifying the dose delivered by a single isocenter stereotactic radiosurgery (SRS) treatment of small multiple brain metastases (BM).
methods14 clinical treatment cases were selected with the number of BM ranging from 2 to 11. The dosimetric agreement was investigated between the calculated and the measured dose by an OCTAVIUS 1600SRS array detector in an OCTAVIUS 4D phantom equipped with dedicated SRS top. The cross-calibration procedure deviated from the manufacturer's as it applied field sizes and dose rates corresponding to the volumetric modulated arc therapy segments in each plan.
resultsMeasurements with a plan specific cross-calibration showed mean ± standard deviation (SD) agreement scores for cut-off values 50%, 80%, 95%, of 98.6 ± 1.7%, 96.5 ± 4.6%, 97.3 ± 4.4% for the 6 MV plans respectively, and 98.6 ± 1.5%, 96.6 ± 4.0% 96.4 ± 6.3%, for the 6 MV flattening filter free (FFF) plans respectively. Using the default calibration procedure instead of the plan specific calibration could lead to a combined systematic dose offset of 4.1% for our treatment plans.
conclusionThe 1600SRS detector array with the 4D phantom offers an accurate solution to perform routine quality assurance measurements of single isocenter SRS treatments of multiple BM. This work points out the necessity of an adapted cross-calibration procedure. ADVANCES IN KNOWLEDGE: A dedicated calibration procedure enables accurate dosimetry with the 1600SRS detector for small field single isocenter SRS treatment of multiple brain metastases for a large amount of BM.