Evidence map›Paper›PMID 34109738›Full record

ArticleJournal of applied clinical medical physics2021

Automated procedure for slice thickness verification of computed tomography images: Variations of slice thickness, position from iso-center, and reconstruction filter.

Nani Lasiyah, Choirul Anam, Eko Hidayanto, Geoff Dougherty

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 5 papers.

0numbers the graph read from it
0cells of the map it votes in
5citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

5 citing papers in PubMed.

  1. Article
  2. Article
  3. Article
  4. Article
  5. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

4 authors.

Nani LasiyahDepartment of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Semarang, Indonesia.
Choirul AnamDepartment of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Semarang, Indonesia.
Eko HidayantoDepartment of Physics, Faculty of Sciences and Mathematics, Diponegoro University, Semarang, Indonesia.
Geoff DoughertyDepartment of Applied Physics and Medical Imaging, California State University Channel Islands, Camarillo, CA, USA.

Funding

World Class Research University 118-08/UN7.6.1/PP/2021
6 · The paper itself

Abstract

purposeThe purpose of this study is to automate the slice thickness verification on the AAPM CT performance phantom and validate it for variations of slice thickness, position from iso-center, and reconstruction filter.

methodsAn automatic procedure for slice thickness verification on AAPM CT performance phantom was developed using MATLAB R2015b. The stair object image within the phantom was segmented, and the middle stair object was located. Its angle was determined using the Hough transformation, and the image was rotated accordingly. The profile through this object was obtained, and its full-width of half maximum (FWHM) was automatically measured. The FWHM indicated the slice thickness of the image. The automated procedure was applied with variations in three independent parameters, i.e., the slice thickness, the distance from the phantom to the iso-center, and the reconstruction filter. The automated results were compared to manual measurements made using electronic calipers.

resultsThe differences of the automated results from the nominal slice thicknesses were within 1.0 mm. The automated results are comparable to those from manual approach (i.e., the difference of both is within 12%). The automatic procedure accurately obtained slice thickness even when the phantom was moved from the iso-center position by up to 4 cm above and 4 cm below the iso-center. The automated results were similar (to within 0.1 mm) for various reconstruction filters.

conclusionsWe successfully developed an automated procedure of slice thickness verification and confirmed that the automated procedure provided accurate results. It provided an easy and effective method of determining slice thickness.

Indexed as

Tomography, X-Ray ComputedHumansPhantoms, ImagingAAPM CT performance phantomCT scannerCT scanner performance evaluationslice sensitivity profileslice thickness

Identifiers

PMID34109738
PMCPMC8292687