Comparison of keratometry data using handheld and table-mounted instruments in healthy adults.

Hagar Olshaker, Ortal Buhbut, Asaf Achiron, Gad Dotan, Issac Levy, Raimo Tuuminen, Assaf Gershoni, Irit Bahar, Uri Elbaz

Abstract read
PubMed Publisher
In one paragraph

Article in International ophthalmology, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Not yet cited in PubMed.

0numbers the graph read from it
0cells of the map it votes in
0citing papers in PubMed
–field-weighted citation impact, top 90% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

0 citing papers in PubMed, 0 citations in OpenAlex.

No citing paper in PubMed yet.

4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

9 authors at 3 institutions in 2 countries.

Hagar OlshakerSackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel.
Ortal BuhbutDepartment of Ophthalmology, The Faculty of Health Sciences, Soroka University Medical Center, Ben Gurion University of the Negev, Beer-Sheva, Israel.
Asaf AchironSackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel.
Gad DotanSackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel.
Issac LevySackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel.
Raimo TuuminenDepartment of Ophthalmology, Kymenlaakso Central Hospital, Kotka, Finland.
Assaf GershoniSackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel.ORCID http://orcid.org/0000-0003-4395-5555
Irit BaharSackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel.
Uri ElbazSackler Faculty of Medicine, Tel Aviv University, Tel Aviv, Israel. urielbaz@gmail.com.
Tel Aviv University · ILBen-Gurion University of the Negev · ILKymenlaakson keskussairaala · FI

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

purposeTo compare keratometry data between the handheld Retinomax K-plus 3 and the table-mounted IOLMaster 700.

methodsHealthy adult volunteers were prospectively recruited to the study. All participants underwent 3 consecutive keratometry measurements using the Retinomax K-plus 3 and a single biometry assessment using the IOLMaster 700. Differences between the Retinomax K-plus 3 and the IOLMaster 700 were assessed using Wilcoxon test for paired samples, Spearman correlation, Bland-Altman and mountain plots.

resultTwenty-eight healthy subjects with a median age of 37 years (interquartile range (IQR) 28-44 years) were included in the study. The median mean keratometry (mean K) reading was higher using the Retinomax K-plus 3 (44.04D; IQR 42.96-45.61D) compared to the IOLMaster 700 (43.78D; IQR 43.22-44.90D, p < 0.01), with a mean difference of 0.18D (95% confidence interval (CI) 0.11-0.23D). Mean K readings were highly correlated between the 2 devices (r = 0.995, p < 0.01). Bland-Altman plots showed 95% limits of agreement between -0.14D and 0.49D. Frequency histogram of mean K reading differences between the Retinomax K-plus 3 and the IOLMaster 700 showed that 56% of cases were between ± 0.2D, 93% of cases were between ± 0.4D and all cases were between ± 0.5D. Mean corneal astigmatism measurement was higher using the Retinomax K-plus 3 (1.01 ± 0.40D) compared to the IOLMaster 700 (0.77 ± 0.36D), with a mean difference of 0.23 ± 0.37D (p < 0.01) between the devices.

conclusionsA good agreement exists between the Retinomax K-plus 3 and the IOLMaster 700 regarding keratometry readings. This enables cataract surgeons to safely use the Retinomax K-plus 3 device when indicated.

Indexed as

AstigmatismCataractAdultBiometryCorneaHumansReproducibility of Results