Evidence map›Paper›PMID 34095419›Full record

ArticlePractical laboratory medicine2021

Role of POC INR in the early stage of diagnosis of coagulopathy.

Florian J Raimann, Marie-Louise Lindner, Christoph Martin, Lukas Jennewein, Thomas Lustenberger, Florian Piekarski, Kai Zacharowski, Christian F Weber

Open access · goldAbstract read
In one paragraph

Article in Practical laboratory medicine, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
0.3field-weighted citation impact, top 44% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed, 3 citations in OpenAlex.

  1. Article
  2. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

8 authors at 3 institutions in 1 country.

Florian J RaimannDepartment of Anaesthesiology, Intensive Care Medicine and Pain Therapy, University Hospital Frankfurt, Goethe University Frankfurt, Germany.
Marie-Louise LindnerDepartment of Anaesthesiology, Intensive Care Medicine and Emergency Medicine, Asklepios Clinics Hamburg, AK Wandsbek, Alphonsstraße 14, 22043, Hamburg, Germany.
Christoph MartinDepartment of Anaesthesiology, Intensive Care Medicine and Pain Therapy, University Hospital Frankfurt, Goethe University Frankfurt, Germany.
Lukas JenneweinDepartment of Gynecology and Obstetrics, University Hospital Frankfurt, Goethe University Frankfurt, Germany.
Thomas LustenbergerDepartment of Trauma, Hand and Reconstructive Surgery, University Hospital Frankfurt, Goethe University Frankfurt, Germany.
Florian PiekarskiDepartment of Anaesthesiology, Intensive Care Medicine and Pain Therapy, University Hospital Frankfurt, Goethe University Frankfurt, Germany.
Kai ZacharowskiDepartment of Anaesthesiology, Intensive Care Medicine and Pain Therapy, University Hospital Frankfurt, Goethe University Frankfurt, Germany.
Christian F WeberDepartment of Anaesthesiology, Intensive Care Medicine and Emergency Medicine, Asklepios Clinics Hamburg, AK Wandsbek, Alphonsstraße 14, 22043, Hamburg, Germany.
University Hospital Frankfurt · DEAsklepios · DEGoethe University Frankfurt · DE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundAcute bleeding requires fast and targeted therapy. Therefore, knowledge of the patient's potential to form a clot is crucial. Point-of-care testing (POCT) provides fast and reliable information on coagulation. Structural circumstances, such as person-bound sample transport, can prolong the reporting of the results. The aim of the present study was to investigate the diagnostic quality and accuracy between POCT INR diagnostics and standard laboratory analysis (SLA) as well as the time advantage between a pneumatic tube and a personal-based transport system.

methodsTwo groups of haemorrhagic patients (EG: emergency department; OG: delivery room; each n = 12) were examined in the context of bleeding emergencies using POCT and SLA. Samples were transported via a pneumatic tube system or by a personal transport service.

resultsINR results between POCT and SLA showed a high and significant correlation (EG: p < 0.001; OG: p < 0.001). POCT results were reported significantly more quickly (EG: 1.1 vs. 39.6 min; OG: 2.0 vs. 75.0 min; p < 0.001) and required less time for analysis (EG: 0.3 vs. 24.0 min; OG: 0.5 vs. 45.0 min; p < 0.001) compared to SLA. The time for transportation with the pneumatic tube was significantly shorter (8.0 vs. 18.5 min; p < 0.001) than with the personal-based transport system.

conclusionThe results of the present study suggest that POCT may be a suitable method for the emergency diagnosis and may be used as prognostic diagnostic elements in haemotherapy algorithms to initiate targeted haemotherapy at an early point in time.

Indexed as

BleedingINRInternational normalized ratioPoint-of-care testing