ArticleMagnetic resonance in medicine2021
Radiofrequency-induced heating of broken and abandoned implant leads during magnetic resonance examinations.
Article in Magnetic resonance in medicine, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 10 papers, 1 of them a synthesis that pooled it.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
10 citing papers in PubMed, 1 synthesis or guideline pooled it, 16 citations in OpenAlex.
- SCMR expert consensus statement for cardiovascular magnetic resonance of patients with a cardiac implantable electronic device.Journal of cardiovascular magnetic resonance : official journal of the Society for Cardiovascular Magnetic Resonance · 2024Guideline
- Assessing radiofrequency safety of active implants by measuring induced radiofrequency currents using MRI.Magnetic resonance in medicine · 2026Article
- ESR Essentials: basic physics of MR safety-practice recommendations by the European Society for Magnetic Resonance in Medicine and Biology.European radiology · 2025Review
- Device-Related Complications Associated with Cylindrical Lead Spinal Cord Stimulator Implants: A Comprehensive Review.Current pain and headache reports · 2024Review
- MRI in Patients with Cardiovascular Implantable Electronic Devices and Fractured or Abandoned Leads.Radiology. Cardiothoracic imaging · 2024Article
- A new method to improve RF safety of implantable medical devices using inductive coupling at 3.0 T MRI.Magma (New York, N.Y.) · 2023Article
- Implants talk to each-other: RF heating changes when two DBS leads are present simultaneously during MRI.Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2023Article
- EHRA consensus on prevention and management of interference due to medical procedures in patients with cardiac implantable electronic devices.Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2022Article
- Magnetic resonance imaging interactions with a sacral neuromodulation system.Neurourology and urodynamics · 2021Article
- Radiofrequency-induced heating of broken and abandoned implant leads during magnetic resonance examinations.Magnetic resonance in medicine · 2021Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
5 authors at 4 institutions in 3 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
purposeThe risks of RF-induced heating of active implantable medical device (AIMD) leads during MR examinations must be well understood and realistically assessed. In this study, we evaluate the potential additional risks of broken and abandoned (cut) leads.
methodsFirst, we defined a generic AIMD with a metallic implantable pulse generator (IPG) and a 100-cm long lead containing 1 or 2 wires. Next, we numerically estimated the deposited in vitro lead-tip power for an intact lead, as well as with wire breaks placed at 10 cm intervals. We studied the effect of the break size (wire gap width), as well as the presence of an intact wire parallel to the broken wire, and experimentally validated the numeric results for the configurations with maximum deposited in vitro lead-tip power. Finally, we performed a Tier 3 assessment of the deposited in vivo lead-tip power for the intact and broken lead in 4 high resolution virtual population anatomic models for over 54,000 MR examination scenarios.
resultsThe enhancement of the deposited lead-tip power for the broken leads, compared to the intact lead, reached 30-fold in isoelectric exposure, and 16-fold in realistic clinical exposures. The presence of a nearby intact wire, or even a nearby broken wire, reduced this enhancement factor to <7-fold over the intact lead.
conclusionBroken and abandoned leads can pose increased risk of RF-induced lead-tip heating to patients undergoing MR examinations. The potential enhancement of deposited in vivo lead-tip power depends on location and type of the wire break, lead design, and clinical routing of the lead, and should be carefully considered when performing risk assessment for MR examinations and MR conditional labeling.