Validation of virtual water phantom software for pre-treatment verification of single-isocenter multiple-target stereotactic radiosurgery.

Juan-Francisco Calvo-Ortega, Peter B Greer, Marcelino Hermida-López, Sandra Moragues-Femenía, Coral Laosa-Bello, Joan Casals-Farran

Abstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.

0numbers the graph read from it
0cells of the map it votes in
2citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

2 citing papers in PubMed.

  1. Article
  2. Verification of stereotactic radiosurgery plans for multiple brain metastases using a virtual phantom-based procedure.Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation Oncology · 2022
    Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

6 authors.

Juan-Francisco Calvo-OrtegaServicio de Oncología Radioterápica, Hospital Quirónsalud, Barcelona, Spain.
Peter B GreerDepartment of Radiation Oncology, Calvary Mater Newcastle Hospital, Newcastle, NSW, 2298, Australia.
Marcelino Hermida-LópezServei de Física i Protecció Radiològica, Hospital Vall d'Hebron, Barcelona, Spain.
Sandra Moragues-FemeníaServicio de Oncología Radioterápica, Hospital Quirónsalud, Barcelona, Spain.
Coral Laosa-BelloServicio de Oncología Radioterápica, Hospital Quirónsalud, Barcelona, Spain.
Joan Casals-FarranServicio de Oncología Radioterápica, Hospital Quirónsalud, Barcelona, Spain.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

The aim of this study was to benchmark the accuracy of the VIrtual Phantom Epid dose Reconstruction (VIPER) software for pre-treatment dosimetric verification of multiple-target stereotactic radiosurgery (SRS). VIPER is an EPID-based method to reconstruct a 3D dose distribution in a virtual phantom from in-air portal images. Validation of the VIPER dose calculation was assessed using several MLC-defined fields for a 6 MV photon beam. Central axis percent depth doses (PDDs) and output factors were measured with an ionization chamber in a water tank, while dose planes at a depth of 10 cm in a solid flat phantom were acquired with radiochromic films. The accuracy of VIPER for multiple-target SRS plan verification was benchmarked against Monte Carlo simulations. Eighteen multiple-target SRS plans designed with the Eclipse treatment planning system were mapped to a cylindrical water phantom. For each plan, the 3D dose distribution reconstructed by VIPER within the phantom was compared with the Monte Carlo simulation, using a 3D gamma analysis. Dose differences (VIPER vs. measurements) generally within 2% were found for the MLC-defined fields, while film dosimetry revealed gamma passing rates (GPRs) ≥95% for a 3%/1 mm criteria. For the 18 multiple-target SRS plans, average 3D GPRs greater than 93% and 98% for the 3%/2 mm and 5%/2 mm criteria, respectively. Our results validate the use of VIPER as a dosimetric verification tool for pre-treatment QA of single-isocenter multiple-target SRS plans. The method requires no setup time on the linac and results in an accurate 3D characterization of the delivered dose.

Indexed as

RadiosurgeryRadiotherapy, Intensity-ModulatedHumansPhantoms, ImagingRadiotherapy DosageRadiotherapy Planning, Computer-AssistedSoftwareWaterWatermultiple-targetSRSvirtual phantom