Comparison of different preparation techniques of dried blood spot quality controls in newborn screening for congenital adrenal hyperplasia.

Nóra Grecsó, Anita Zádori, Ákos Baráth, Zsolt Galla, Gábor Rácz, Csaba Bereczki, Péter Monostori

Open access · goldAbstract read
In one paragraph

Article in PloS one, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 3 papers.

0numbers the graph read from it
0cells of the map it votes in
3citing papers in PubMed
0.3field-weighted citation impact, top 46% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

3 citing papers in PubMed, 4 citations in OpenAlex.

  1. Review
  2. Article
  3. Review
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 1 institution in 1 country.

Nóra GrecsóMetabolic and Newborn Screening Laboratory, Department of Pediatrics, University of Szeged, Szeged, Hungary.ORCID 0000-0002-3355-9113
Anita ZádoriMetabolic and Newborn Screening Laboratory, Department of Pediatrics, University of Szeged, Szeged, Hungary.
Ákos BaráthMetabolic and Newborn Screening Laboratory, Department of Pediatrics, University of Szeged, Szeged, Hungary.
Zsolt GallaMetabolic and Newborn Screening Laboratory, Department of Pediatrics, University of Szeged, Szeged, Hungary.
Gábor RáczMetabolic and Newborn Screening Laboratory, Department of Pediatrics, University of Szeged, Szeged, Hungary.
Csaba BereczkiMetabolic and Newborn Screening Laboratory, Department of Pediatrics, University of Szeged, Szeged, Hungary.
Péter MonostoriMetabolic and Newborn Screening Laboratory, Department of Pediatrics, University of Szeged, Szeged, Hungary.ORCID 0000-0003-3591-6054
University of Szeged · HU

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

In newborn screening, samples suspected for congenital adrenal hyperplasia (CAH), a potentially lethal inborn error of steroid biosynthesis, need to be confirmed using liquid chromatography-tandem mass spectrometry. Daily quality controls (QCs) for the 2nd-tier CAH assay are not commercially available and are therefore generally prepared within the laboratory. For the first time, we aimed to compare five different QC preparation approaches used in routine diagnostics for CAH on the concentrations of cortisol, 21-deoxycortisol, 11-deoxycortisol, 4-androstenedione and 17-hydroxyprogesterone in dried blood spots. The techniques from Prep1 to Prep5 were tested at two analyte concentrations by spiking aliquots of a steroid-depleted blood, derived from washed erythrocyte suspension and steroid-depleted serum. The preparation processes differed in the sequence of the preparation steps and whether freeze-thaw cycles were used to facilitate blood homogeneity. The five types of dried blood spot QCs were assayed and quantitated in duplicate on five different days using a single calibration row per day. Inter-assay variations less than 15% and concentrations within ±15% of the nominal values were considered acceptable. Results obtained by means of the four dried blood spot QC preparation techniques (Prep1, Prep2, Prep4 and Prep5) were statistically similar and remained within the ±15% ranges in terms of both reproducibility and nominal values. However, concentration results for Prep3 (spiking prior to three freeze-thaw cycles) were significantly lower than the nominal values in this setting, with differences exceeding the ±15% range in many cases despite acceptable inter-assay variations. These findings have implications for the in-house preparation of QC samples in laboratory developed tests for CAH, including 2nd-tier assays in newborn screening.

Indexed as

17-alpha-HydroxyprogesteroneAdrenal Hyperplasia, CongenitalAndrostenedioneCortodoxoneDried Blood Spot TestingHumansInfant, NewbornNeonatal ScreeningTandem Mass Spectrometry17-alpha-Hydroxyprogesterone21-deoxycortisolAndrostenedioneCortodoxone