A material stress test study on occurrence of leakage and material failure of peritoneal dialysis (PD) catheters.

Matthias Klingele, Martin Carstens, Lea Baerens, Matthias W Laschke, Wolfgang Metzger, Danilo Fliser, Clemens M Meier

Open access · goldAbstract read
In one paragraph

Article in Scientific reports, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed, 1 pooled it
0.3field-weighted citation impact, top 45% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 1 synthesis or guideline pooled it, 7 citations in OpenAlex.

  1. Pooled it
  2. Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 2 institutions in 1 country.

Matthias KlingeleDepartment of Nephrology, Hochtaunuskliniken, Zeppelinstrasse 20, 61352, Bad Homburg, Germany. klingelemat@yahoo.fr.
Martin CarstensDepartment of Internal Medicine, Nephrology and Hypertension, Saarland University Medical Centre, Homburg, Saar, Germany.
Lea BaerensDepartment of Internal Medicine, Nephrology and Hypertension, Saarland University Medical Centre, Homburg, Saar, Germany.
Matthias W LaschkeInstitute for Clinical and Experimental Surgery, Saarland University Medical Centre, Homburg, Saar, Germany.
Wolfgang MetzgerDepartment of Trauma, Hand and Reconstructive Surgery, Saarland University Medical Centre, Homburg, Saar, Germany.
Danilo FliserDepartment of Internal Medicine, Nephrology and Hypertension, Saarland University Medical Centre, Homburg, Saar, Germany.
Clemens M MeierDepartment of General, Visceral, Vascular and Pediatric Surgery, Saarland University Medical Centre, Homburg, Saar, Germany.
Saarland University · DEAccadis Hochschule Bad Homburg · DE

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

Peritonitis is a common complication of peritoneal dialysis (PD). Our root cause analysis allowed to attribute some cases to leakage of the PD catheter. Accordingly, a clinically based stress test study on potential material damage issues of PD catheters was performed, focusing on material damage caused by cleaning, de- and attachment procedures during dialysate changes and on the individual storage methods of PD catheters between dialysate changes. PD catheters were exposed to both chemical stress by repeating dialysate-flow and physical stress simulating de- and connecting, fixation, pressure, flexing, folding etc.-simulating standard clinical daily routine of 8-10 years PD catheter usage. Potentially by normal usage caused damages should be then detected by intraluminal pressure, light- and electron microscopy. The multi-step visual control showed no obvious damages on PD catheters nor any leakage or barrier indulgence. Our tests simulating daily routine usage of PD catheters for several years could not detect any material defects under chemical or physical stress. Hence, we presume that most PD catheter damages, as identified cause for peritonitis in some of our patients, may be due to accidental, unnoticed external damage (e.g. through scissors, while changing dressings) or neglecting PD catheter handling specifications.

Indexed as

CathetersCatheters, IndwellingEquipment FailureHumansMechanical TestsPeritoneal DialysisPeritoneal Dialysis, Continuous AmbulatoryPeritonitisStress, Mechanical

Identifiers

PMID33986385
PMCPMC8119981
OpenAlexW3160718512

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.