Evidence map›Paper›PMID 33975591›Full record

ArticleBiomedical engineering online2021

Understanding the influence of left ventricular assist device inflow cannula alignment and the risk of intraventricular thrombosis.

Michael Neidlin, Sam Liao, Zhiyong Li, Benjamin Simpson, David M Kaye, Ulrich Steinseifer, Shaun Gregory

Open access · goldAbstract read
In one paragraph

Article in Biomedical engineering online, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 4 papers.

0numbers the graph read from it
0cells of the map it votes in
4citing papers in PubMed
1.6field-weighted citation impact, top 18% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

4 citing papers in PubMed, 24 citations in OpenAlex.

  1. Article
  2. Relapsing low-flow alarms due to suboptimal alignment of the left ventricular assist device inflow cannula.European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery · 2022
    Article
  3. Article
  4. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 5 institutions in 3 countries.

Michael NeidlinDepartment of Cardiovascular Engineering, Institute of Applied Medical Engineering, Helmholtz Institute, RWTH Aachen University, Pauwelsstraße 20, 52074, Aachen, Germany. neidlin@ame.rwth-aachen.de.ORCID http://orcid.org/0000-0002-5582-6905
Sam LiaoDepartment of Cardiovascular Engineering, Institute of Applied Medical Engineering, Helmholtz Institute, RWTH Aachen University, Pauwelsstraße 20, 52074, Aachen, Germany.
Zhiyong LiInstitute of Health and Biomedical Innovation (IHBI), Queensland University of Technology (QUT), Kelvin Grove, QLD, 4059, Australia.
Benjamin SimpsonDepartment of Engineering, Nottingham Trent University, Clifton Lane, Nottingham, NG11 8NS, UK.
David M KayeCardio-Respiratory Engineering and Technology Laboratory (CREATElab), Baker Heart and Diabetes Institute, Melbourne, VIC, 3004, Australia.
Ulrich SteinseiferDepartment of Cardiovascular Engineering, Institute of Applied Medical Engineering, Helmholtz Institute, RWTH Aachen University, Pauwelsstraße 20, 52074, Aachen, Germany.
Shaun GregoryDepartment of Mechanical and Aerospace Engineering, Monash University, Clayton, VIC, 3800, Australia.
Baker Heart and Diabetes Institute · AUMonash University · AUNottingham Trent University · GBQueensland University of Technology · AURWTH Aachen University · DE

Funding

National Health and Medical Research Council of Australia (AU) APP1079421National Heart Foundation of Australia 102062National Heart Foundation of Australia 102499The Prince Charles Hospital Foundation (AU) MS2013-29The Prince Charles Hospital Foundation (AU) PRO2014-08The Prince Charles Hospital Foundation (AU) TM2017-04
6 · The paper itself

Abstract

backgroundAdverse neurological events associated with left ventricular assist devices (LVADs) have been suspected to be related to thrombosis. This study aimed to understand the risks of thrombosis with variations in the implanted device orientation. A severely dilated pulsatile patient-specific left ventricle, modelled with computational fluid dynamics, was utilised to identify the risk of thrombosis for five cannulation angles. With respect to the inflow cannula axis directed towards the mitral valve, the other angles were 25° and 20° towards the septum and 20° and 30° towards the free wall.

resultsInflow cannula angulation towards the free wall resulted in longer blood residence time within the ventricle, slower ventricular washout and reduced pulsatility indices along the septal wall. Based on the model, the ideal inflow cannula alignment to reduce the risk of thrombosis was angulation towards the mitral valve and up to parallel to the septum, avoiding the premature clearance of incoming blood.

conclusionsThis study indicates the potential effects of inflow cannulation angles and may guide optimised implantation configurations; however, the ideal approach will be influenced by other patient factors and is suspected to change over the course of support.

Indexed as

CannulaHeart-Assist DevicesHeart VentriclesThrombosisCatheterizationHumansHydrodynamicsModels, CardiovascularRiskCannula positionCardiomyopathyCardiovascular surgeryComputational fluid dynamicsHeart failureLeft ventricular assist deviceThrombosis

Identifiers

PMID33975591
PMCPMC8114696
OpenAlexW3163498425

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.