ArticleMagnetic resonance in medicine2021
Vertical open-bore MRI scanners generate significantly less radiofrequency heating around implanted leads: A study of deep brain stimulation implants in 1.2T OASIS scanners versus 1.5T horizontal systems.
Article in Magnetic resonance in medicine, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 25 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
25 citing papers in PubMed, 39 citations in OpenAlex.
- Comparative study of radiofrequency heating in deep brain stimulation devices during MRI at 1.5 T and 0.55 T: Challenging the assumption of safety at low field strengths.Magnetic resonance in medicine · 2025Article
- Simulated radiation levels and patterns of MRI without a Faraday shielded room.Magnetic resonance in medicine · 2025Article
- Control of radiated emissions during MRI using parallel transmit: a path to eliminating the Faraday shielded room for 0.5T imaging.Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2025Article
- Enhancing fine-tuning efficiency and design optimization of an eight-channel 3T transmit array via equivalent circuit modeling and Eigenmode analysis.Medical physics · 2025Article
- A deep brain stimulation-conditioned RF coil for 3T MRI.Magnetic resonance in medicine · 2025Article
- Dual birdcage coils to control radiation from MRI systems operated without a Faraday cage.Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2024Article
- Effect of surgical modification of deep brain stimulation lead trajectories on radiofrequency heating during MRI at 3T: from phantom experiments to clinical implementation.Journal of neurosurgery · 2024Article
- Comparison of tight-fitting 7T parallel-transmit head array designs using excitation uniformity and local specific absorption rate metrics.Magnetic resonance in medicine · 2024Article
- Modifying the trajectory of epicardial leads can substantially reduce MRI-induced RF heating in pediatric patients with a cardiac implantable electronic device at 1.5T.Magnetic resonance in medicine · 2023Article
- Age Matters: A Comparative Study of RF Heating of Epicardial and Endocardial Electronic Devices in Pediatric and Adult Phantoms during Cardiothoracic MRI.Diagnostics (Basel, Switzerland) · 2023Article
- RF-induced heating of capped and uncapped abandoned epicardial leads during MRI at 1.5 T and 3 T.Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2023Article
- RF-induced heating of interventional devices at 23.66 MHz.Magma (New York, N.Y.) · 2023Article
- Implants talk to each-other: RF heating changes when two DBS leads are present simultaneously during MRI.Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2023Article
- Rapid prediction of MRI-induced RF heating of active implantable medical devices using machine learning.Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2023Article
- Low-field MRI's Spark on Implant Safety: A Closer Look at Radiofrequency Heating.Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2023Article
- Optimizing the trajectory of deep brain stimulation leads reduces RF heating during MRI at 3 T: Characteristics and clinical translation.Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2023Article
- Multi-Segment Leads To Reduce RF Heating in MRI: A Computational Evaluation at 1.5T and 3T.Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2023Article
- Application of Machine learning to predict RF heating of cardiac leads during magnetic resonance imaging at 1.5 T and 3 T: A simulation study.Journal of magnetic resonance (San Diego, Calif. : 1997) · 2023Article
- Article
- A workflow for predicting temperature increase at the electrical contacts of deep brain stimulation electrodes undergoing MRI.Magnetic resonance in medicine · 2022Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
11 authors at 5 institutions in 2 countries.
Funding
Abstract
purposePatients with active implants such as deep brain stimulation (DBS) devices are often denied access to MRI due to safety concerns associated with the radiofrequency (RF) heating of their electrodes. The majority of studies on RF heating of conductive implants have been performed in horizontal close-bore MRI scanners. Vertical MRI scanners which have a 90° rotated transmit coil generate fundamentally different electric and magnetic field distributions, yet very little is known about RF heating of implants in this class of scanners. We performed numerical simulations as well as phantom experiments to compare RF heating of DBS implants in a 1.2T vertical scanner (OASIS, Hitachi) compared to a 1.5T horizontal scanner (Aera, Siemens).
methodsSimulations were performed on 90 lead models created from post-operative CT images of patients with DBS implants. Experiments were performed with wires and commercial DBS devices implanted in an anthropomorphic phantom.
resultsWe found significant reduction of 0.1 g-averaged specific absorption rate (30-fold, P < 1 × 10
conclusionVertical MRI scanners appear to generate lower RF heating around DBS leads, providing potentially heightened safety or the flexibility to use sequences with higher power levels than on conventional systems.