Estimating the "Pull" on a Pullthrough Wire: A Pilot Study.
Article in EJVES vascular forum, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 2 papers.
What it found
Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.
The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.
The trial behind it
Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.
Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.
Who cites it
2 citing papers in PubMed, 3 citations in OpenAlex.
- Quantifying the Functional Stiffness of Pullthrough Wires Used for Endovascular Aneurysm Repairs Using Comparative Tension Dynamometry.EJVES vascular forum · 2022Article
- Are All Wires Created the Same? A Quality Assurance Study of the Stiffness of Wires Typically Employed During Endovascular Surgery Using Tension Dynamometry.EJVES vascular forum · 2021Article
Corrections and comments
PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.
Authors and funding
3 authors at 2 institutions in 2 countries.
Funding
No grant is acknowledged in the PubMed record.
Abstract
objectivePullthrough/body floss wires are used to track endovascular devices across tortuous aorto-iliac anatomy encountered during endovascular repair of abdominal or thoracic aortic aneurysms. The tension imparted on such wires is arbitrary and has never been quantified. This pilot study attempted to quantify the tension used to stiffen the floppy hydrophilic wires typically used in such a scenario.
methodsTwo linked experiments were undertaken, the first by tasking 13 blinded vascular surgeons (eight male, five female; mean age 36 ± 11 years, including nine trainees) with pulling a long floppy hydrophilic wire (Radifocus Guidewire M Stiff, Terumo UK, Bagshot, Surrey, UK) attached at the other end to a horizontally configured industrial scale (HDN-N Hanging Scale, Kern & Sohn GmbH, Balingen, Germany), to simulate what they individually felt was an "appropriate" tension; the second by using the derived average tensioning force to set up a pullthrough wire within a rigid life like aorto-iliac model to assess whether a test device (16F Sentrant Introducer Sheath, Medtronic Limited, Watford, UK) could be delivered over such a tensioned wire in both brachiofemoral and femorofemoral configurations.
resultsThe mean tension exerted by the group on the wire was 38.3 ± 14.8 N (equivalent to 3.9 kgf). Pullthrough wire tensioning was undertaken by fixing one end and applying a 3.9 kg weight at the other. The test device was successfully deployed into the infrarenal aortic position and also across the aortic bifurcation, via brachiofemoral and femorofemoral pullthrough configurations, respectively.
conclusionSuccessful test device deliveries suggest that a minimum tension equivalent to almost 4 kgf applied to a floppy wire can provide "stiffeningˮ to allow device tracking across tortuous aorto-iliac anatomy. More studies are needed to ascertain whether lower tensions can be applied; these results may help provide a platform for other such studies depending on configuration, aortic geometry, and device or wire/tension characteristics.