Evidence map›Paper›PMID 33869708›Full record

ArticlePractical laboratory medicine2021

Six Sigma performance of quality indicators in total testing process of point-of-care glucose measurement: A two-year review.

Anne Vincent, Donnah Pocius, Yun Huang

Abstract read
In one paragraph

Article in Practical laboratory medicine, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 1 paper.

0numbers the graph read from it
0cells of the map it votes in
1citing papers in PubMed
–field-weighted citation impact
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

1 citing paper in PubMed.

  1. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

3 authors.

Anne VincentKingston General Hospital, 76 Stuart Street, Kingston, ON, Canada.
Donnah PociusKingston General Hospital, 76 Stuart Street, Kingston, ON, Canada.
Yun HuangKingston General Hospital, 76 Stuart Street, Kingston, ON, Canada.

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

objectivesThe error rate in the total testing process (TTP) of point-of-care (POC) glucose measurement remains high although a total quality management system has been applied. Quality indicators (QIs) in the TTP of glucose meter were established via risk assessment. Their two-year Six Sigma values were reviewed for quality improvement.

designThe TTP of POC glucose measurement was mapped to identify risks in key steps. The risks were assessed for their frequency and severity of impact on patient safety. Whenever possible, measurable data from the data management system and other sources was collected to establish QIs for risk monitoring. Average Six Sigma value of each QI in the last two years was calculated for acceptance and for determining corrective action.

results29 risks were identified in eight key steps of the TTP. Eight QIs were established for monitoring six risks and three QIs for two accepted risks were established for improving operator testing skill. The QIs had a good coverage to key steps. Two, five and four QIs showed Six Sigma values <3, 3-4 and >4 respectively. Six Sigma values of two QIs related to quality control (QC) testing were improved by using meters with accurate QC sample loading.

conclusionsThe establishment of QIs for glucose measurement by risk assessment with measurable data from the data management system and on Six sigma scale was effective, efficient, and manageable. Most of QIs' Six Sigma values were between 3 and 5, which could be improved by using upgraded meters.

Indexed as

Glucose meterPoint-of-care testingQuality indicatorRisk assessmentSix sigma

Identifiers

PMID33869708
PMCPMC8042413

What OpenQuestion holds

Textmetadata
LicenceCC BY-NC-ND
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.