ArticleJournal of applied clinical medical physics2021

Commissioning of optical surface imaging systems for cranial frameless stereotactic radiosurgery.

Lei Zhang, Sarath Vijayan, Sheng Huang, Yulin Song, Tianfang Li, Xiang Li, Elizabeth Hipp, Maria F Chan, Hsiang-Chi Kuo, Xiaoli Tang and 5 more

Open access · goldAbstract read
In one paragraph

Article in Journal of applied clinical medical physics, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 11 papers.

0numbers the graph read from it
0cells of the map it votes in
11citing papers in PubMed
2.5field-weighted citation impact, top 11% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

11 citing papers in PubMed, 11 citations in OpenAlex.

  1. Article
  2. Tattoo-less chest wall and regional nodal irradiation using surface imaging.Technical innovations & patient support in radiation oncology · 2025
    Article
  3. Article
  4. Article
  5. Commissioning and clinical evaluation of the IDENTIFYJournal of applied clinical medical physics · 2023
    Article
  6. Review
  7. Analysis of a surface imaging system using a six degree-of-freedom couch.Journal of applied clinical medical physics · 2022
    Article
  8. Article
  9. Article
  10. Article
  11. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

15 authors at 1 institution in 1 country.

Lei ZhangDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, USA.ORCID https://orcid.org/0000-0001-5369-5756
Sarath VijayanDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, USA.
Sheng HuangDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, USA.
Yulin SongDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, Bergen, NJ, USA.
Tianfang LiDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, Commack, NY, USA.
Xiang LiDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, Nassau, NY, USA.
Elizabeth HippDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, Monmouth, NJ, USA.
Maria F ChanDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, Baskin Ridge, NJ, USA.ORCID https://orcid.org/0000-0002-9140-8604
Hsiang-Chi KuoDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, Norwalk, CT, USA.
Xiaoli TangDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, Westchester, NY, USA.
Grace TangDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, USA.
Seng Boh LimDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, USA.ORCID https://orcid.org/0000-0002-4054-8789
Dale Michael LovelockDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, USA.ORCID https://orcid.org/0000-0002-9371-3573
Ase BallangrudDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, USA.
Guang LiDepartment of Medical Physics, Memorial Sloan Kettering Cancer Center, New York, NY, USA.ORCID https://orcid.org/0000-0002-9022-2883
Memorial Sloan Kettering Cancer Center · US

Funding

X-RAY CRYSTALLOGRAPHYP30CA008748 · NCI · SLOAN-KETTERING INSTITUTE FOR CANCER RES · PI Nicholas D Socci · 1985 to 2026
$347.4M
NCI NIH HHS P30 CA008748NIH HHS P30 CA008748
6 · The paper itself

Abstract

purposeThis study aimed to evaluate and compare different system calibration methods from a large cohort of systems to establish a commissioning procedure for surface-guided frameless cranial stereotactic radiosurgery (SRS) with intrafractional motion monitoring and gating. Using optical surface imaging (OSI) to guide non-coplanar SRS treatments, the determination of OSI couch-angle dependency, baseline drift, and gated-delivered-dose equivalency are essential.

methodsEleven trained physicists evaluated 17 OSI systems at nine clinical centers within our institution. Three calibration methods were examined, including 1-level (2D), 2-level plate (3D) calibration for both surface image reconstruction and isocenter determination, and cube phantom calibration to assess OSI-megavoltage (MV) isocenter concordance. After each calibration, a couch-angle dependency error was measured as the maximum registration error within the couch rotation range. A head phantom was immobilized on the treatment couch and the isocenter was set in the middle of the brain, marked with the room lasers. An on-site reference image was acquired at couch zero, the facial region of interest (ROI) was defined, and static verification images were captured every 10° for 0°-90° and 360°-270°. The baseline drift was assessed with real-time monitoring of the motionless phantom over 20 min. The gated-delivered-dose equivalency was assessed using the electron portal imaging device and gamma test (1%/1mm) in reference to non-gated delivery.

resultsThe maximum couch-angle dependency error occurs in longitudinal and lateral directions and is reduced significantly (P < 0.05) from 1-level (1.3 ± 0.4 mm) to 2-level (0.8 ± 0.3 mm) calibration. The MV cube calibration does not further reduce the couch-angle dependency error (0.8 ± 0.2 mm) on average. The baseline drift error plateaus at 0.3 ± 0.1 mm after 10 min. The gated-delivered-dose equivalency has a >98% gamma-test passing rate.

conclusionA commissioning method is recommended using the 3D plate calibration, which is verified by radiation isocenter and validated with couch-angle dependency, baseline drift, and gated-delivered-dose equivalency tests. This method characterizes OSI uncertainties, ensuring motion-monitoring accuracy for SRS treatments.

Indexed as

RadiosurgeryHumansPatient PositioningPhantoms, ImagingRadiotherapy DosageSkullcalibration and commissioningcouch-angle dependencyframeless stereotactic radiosurgery (SRS)optical surface imaging (OSI)surface-guided frameless radiosurgery