Measurement of Heart Rate Using the Polar OH1 and Fitbit Charge 3 Wearable Devices in Healthy Adults During Light, Moderate, Vigorous, and Sprint-Based Exercise: Validation Study.

David Joseph Muggeridge, Kirsty Hickson, Aimie Victoria Davies, Oonagh M Giggins, Ian L Megson, Trish Gorely, Daniel R Crabtree

Open access · goldAbstract read
In one paragraph

Article in JMIR mHealth and uHealth, 2021. The graph could read no effect estimate from its abstract, so it casts no vote on the map. Cited by 35 papers, 1 of them a synthesis that pooled it.

0numbers the graph read from it
0cells of the map it votes in
35citing papers in PubMed, 1 pooled it
3.8field-weighted citation impact, top 6% of its field
1 · What the graph read from it

What it found

Each row is one number read from the abstract, on the scale the paper reported it, with its interval. Left of the dashed line favours the treatment, right favours the comparator. Under each row is the sentence it came from. New to these charts? A ten-minute tutorial.

The abstract states no effect estimate the extractor could read, or names no intervention and outcome on the map, so this paper lights no cell and moves no belief. It is still indexed, cited and linked below.

2 · The registry

The trial behind it

Trials whose registry record cites this paper, or whose number appears in the abstract. A trial that started after this paper was published is citing it as background, not reporting it.

Neither the registry nor the abstract names a trial number. If this is a trial report, that itself is worth knowing.

3 · Its place in the literature

Who cites it

35 citing papers in PubMed, 1 synthesis or guideline pooled it, 64 citations in OpenAlex.

  1. Pooled it
  2. Trial
  3. Article
  4. Article
  5. Article
  6. Article
  7. Article
  8. Article
  9. Article
  10. Article
  11. Article
  12. Article
  13. Article
  14. Article
  15. Article
  16. Article
  17. Article
  18. Article
  19. Article
  20. Article
4 · The record

Corrections and comments

PubMed lists nothing against this paper. Absence here is not a guarantee, only a check that was made.

5 · Who and what money

Authors and funding

7 authors at 3 institutions in 2 countries.

David Joseph Muggeridge *Edinburgh Napier University, Edinburgh, United Kingdom.ORCID 0000-0003-2630-2384
Kirsty HicksonDivision of Biomedical Sciences, University of the Highlands and Islands, Inverness, United Kingdom.ORCID 0000-0003-2450-5145
Aimie Victoria DaviesDepartment of Nursing and Midwifery, University of the Highlands and Islands, Inverness, United Kingdom.ORCID 0000-0002-5708-1206
Oonagh M GigginsNetwellCASALA, Dundalk Institute of Technology, Dundalk, Ireland.ORCID 0000-0001-7748-5446
Ian L MegsonDivision of Biomedical Sciences, University of the Highlands and Islands, Inverness, United Kingdom.ORCID 0000-0001-8287-2459
Trish GorelyDepartment of Nursing and Midwifery, University of the Highlands and Islands, Inverness, United Kingdom.ORCID 0000-0001-7367-0679
Daniel R Crabtree *Division of Biomedical Sciences, University of the Highlands and Islands, Inverness, United Kingdom.ORCID 0000-0003-2013-9109
University of the Highlands and Islands · GBDundalk Institute of Technology · IEEdinburgh Napier University · GB

Funding

No grant is acknowledged in the PubMed record.

6 · The paper itself

Abstract

backgroundAccurate, continuous heart rate measurements are important for health assessment, physical activity, and sporting performance, and the integration of heart rate measurements into wearable devices has extended its accessibility. Although the use of photoplethysmography technology is not new, the available data relating to the validity of measurement are limited, and the range of activities being performed is often restricted to one exercise domain and/or limited intensities.

objectiveThe primary objective of this study was to assess the validity of the Polar OH1 and Fitbit Charge 3 devices for measuring heart rate during rest, light, moderate, vigorous, and sprint-type exercise.

methodsA total of 20 healthy adults (9 female; height: mean 1.73 [SD 0.1] m; body mass: mean 71.6 [SD 11.0] kg; and age: mean 40 [SD 10] years) volunteered and provided written informed consent to participate in the study consisting of 2 trials. Trial 1 was split into 3 components: 15-minute sedentary activities, 10-minute cycling on a bicycle ergometer, and incremental exercise test to exhaustion on a motorized treadmill (18-42 minutes). Trial 2 was split into 2 components: 4 × 15-second maximal sprints on a cycle ergometer and 4 × 30- to 50-m sprints on a nonmotorized resistance treadmill. Data from the 3 devices were time-aligned, and the validity of Polar OH1 and Fitbit Charge 3 was assessed against Polar H10 (criterion device). Validity was evaluated using the Bland and Altman analysis, Pearson moment correlation coefficient, and mean absolute percentage error.

resultsOverall, there was a very good correlation between the Polar OH1 and Polar H10 devices (r=0.95), with a mean bias of -1 beats·min

conclusionsCompared with our criterion device, Polar OH1 was accurate at assessing heart rate, but the accuracy of Fitbit Charge 3 was generally poor. Polar OH1 performed worse during trial 2 compared with the activities in trial 1, and the validity of the Fitbit Charge 3 device was particularly poor during our cycling exercises.

Indexed as

Fitness TrackersWearable Electronic DevicesAdultExerciseFemaleHeart RateHeart Rate DeterminationHumansexerciseheart ratemobile phonephotoplethysmographyvalidation studywearable electronic devices

Identifiers

PMID33764310
PMCPMC8088863
OpenAlexW3138440227

What OpenQuestion holds

Textmetadata
LicenceCC BY
Read underepoch 390

Registered trials

None linked

Read under generation 80e0d062 · epoch 390. Bibliography from PubMed, PubMed Central and OpenAlex; grants from NIH RePORTER; trial links from ClinicalTrials.gov; estimates, votes and beliefs from the OpenQuestion graph.